Aqua
Fluor
®
U
ser’s Manual
10
On a daily basis, before making measurements with the
Aqua
Fluor
®
, use the solid
secondary standard to verify the
Aqua
Fluor
®
calibration. If the secondary standard value
has changed by more than ± 5% of its assigned value, the
Aqua
Fluor
®
should be
recalibrated.
NOTE: Recalibration with a primary standard if possible is recommended.
2.6 Sample Analysis
1. Insert your sample. The
Aqua
Fluor
®
is designed with “ambient light rejection”. The
black sample compartment cover does NOT need to be closed when reading samples.
NOTE: The orientation and cleanliness of the cuvettes can have an impact on the
accuracy of your results. Refer to Section 4 for more details.
2. Press either <READ> button. The instrument will measure and average the
fluorescence signal for approximately 10 seconds.
3. The reading result will be displayed on the top line of the Home screen.
4.
The top left corner will then display “WAIT” for 5 seconds. Once “WAIT” disappears,
another sample reading can be performed.
2.7 Diagnostic Information
1. Press <DIAG> to access the diagnostic screens.
2. The first screen shows the number of data points available for internal data logging.
3. Press <ENT> to toggle to the %FS (Full Scale) values for the blank (Blk) and standard
(STD) calibration points. This is often referred to as the Standard to Blank Ratio. For
more information please refer to section 2.4.3.
4. Press <ESC> when finished to return to the Home screen.
Summary of Contents for AquaFluor
Page 20: ...AquaFluor User s Manual 20 ...