Appendix A: Specifications
Cuvette information
Cuvette type
Size
Usage
Part No.
Methacrylate
plastic
10 mm Square
(height 44.5 mm)
Required for UV (< 400 nm),
works for all applications.
Do NOT use with solvents
(ie. Acetone)
7000-959
Polystyrene
Plastic
10 mm Square
(height 47.5 mm)
Preferred for Turbidity use, will
NOT work for UV.
Do NOT use with solvents
(ie. Acetone)
7000-957
Borosilicate
glass
12 x 75mm
Round
Required for Extracted
Chlorophyll where solvents are
used. Requires Adapter
PN 8000-932
10-029A
Minicells
–
Borosilicate
Glass
5 x 31mm Flat
bottom
For use with high concentration
samples. Requires Adapter
PN 8000-936 (200 µL capacity)
7000-950
Aqua
Fluor
®
Size
1.75” x 3.5” x 7.25”
(4.45cm x 8.9cm x 18.4cm)
Weight
13.9oz (0.4kg)
Resolution
12 bits
LCD Display
2 x 16 characters
Case
Meets IP 67 Standard; dustproof and waterproof
Temperature
41-104 °F; 5-40 °C
Detector
Photodiode
Calibration Type
Single-point and blank
Error message
Low battery, High blank
Cuvette Type
See below
Warm Up Time
5 seconds
Auto Power Off
After 3 minutes of inactivity