Full Programming Guide
Menu 1 - Adjustable Sensitivity
Green LED on
This menu item is used to set the shock level that will initiate the algorithm into determining
whether a fall has taken place. There are 3 sensitivity settings.
Pressing the Help button will cycle between High/Medium/low levels. After each press a
different tone will sound to indicate the setting selected. When the correct setting is selected,
either press the cancel button to move to Menu 2 or press no further buttons and after 20
seconds the iVi will exit the programming menu.
Option
Sensitivity
Indicator Tone
Note
1
High
High Tone
Very inactive user
2
Medium
Medium Tone
Less active user
3
Low
Low Tone
Default setting
Menu 2 – Equipment Type
Red LED on
Normal Mode
The normal mode is for use with all Tunstall home units
and radio systems. It prevents the ‘Activation Cancelled’
and ‘Not Worn’ alerts from being sent by iVi as these alerts are
not supported by all systems.
Advanced Mode
Must only be used with Lifeline Vi/Vi+ and when ‘Activation
Cancelled’ and ‘Not Worn’ alerts are also required.
Pressing the Help button will cycle between Advanced/Normal mode. After each press a
different tone will sound to indicate the setting selected. When the correct setting is selected,
either press the cancel button to move to Menu 3 or press no further buttons and after 20
seconds the iVi will exit the programming menu.
Option
Setting
Indicator Tone
Note
1
Advanced Mode
High Tone
To be used with Lifeline
Vi/VI+ Only
Activates
cancelled and not worn
alarms
2
Normal Mode
Low Tone
Default setting
Supports Lifeline 400, 4000+,
Connect, , GSM
Lifeline/Caresse, Vi, Vi+,
CareAssist, Telecare
enabled Communicall
systems
Press
Help
button
Press
Help
button