CALIBRATION
Step Calibration (Full calibration Mode)
Edit Values (Adjusting Calibration Values)
Sample(Changing sampling rate)
DpPoint (Deactivated)
Resolution ( Increment or decrement )
1.
2.
3.
4.
5.
Step Calibration Mode is full calibration Mode at 10 multiple points( at every 10 % ) covering
complete instrument range to provide high accuracy of 0.1 % FS throughout the operating range
Use this Mode only when you have calibrated external pressure reference source & hi accuracy
pressure display to calibrate this instrument
Connect the External reference pressure source with calibrated display to the instrument &
following procedure.
Enter the 1. Step calibration mode with help of enter key, After the LCD Display shows the
following prompt
STEP CALIBRATION MODE
Cali.>Step
-------------------------------
Pressure at Zero
+000.000
Adc Count
XXXX
Enter the correct Value of the pressure with help of shift left & Up arrow key