After pressure value shown in zero, Press Enter key to move next 10% point, always
remember to press enter key to savethe value of pressure
After pressure value shown in 10%, Press Enter key to move next 20% point,
Cali.>Step
-------------------------------
Pressure at 10%
+000.030
Adc Count
XXXX
Follow same steps at every 10 % till step calibration is complete till SPAN , a message will appear
that calibration is complete
ACTIVATION OF “EDIT VALUES “IN CALIBRATION MODE
Edit Values correction is used to perform small corrections in calibration values, specific to
the calibration points or defined range
After Entering into setting Mode/Calibration
The calibration mode with help of enter key, After the LCD Display shows the following
prompt