![Tunix TDP-01B Instrument Manual Download Page 12](http://html1.mh-extra.com/html/tunix/tdp-01b/tdp-01b_instrument-manual_1180053012.webp)
After display edit table in display inter Sr no.1 pressure at Zero with help of inter key (Zero key),
Pressure at Zero
-------------------------------
1. Pressure : +000.000
2. Count : 208203
Edit the Pressure Value & ADC count value according to pressure value With help of left arrow (Unit)
and up arrow key, after editing of count value move to edit table with help of exit key (light key).
After Show edit table in display inter Sr no.2 pressure at 10% with help of inter key (Zero key),
Pressure at 10%
-------------------------------
1.Pressure : +000.030
2.Count : 224746
Edit the count value according to pressure value With help of left arrow(Unit) and up arrow key, after
editing of count value move to edit table with help of exit key (light key).