TDP-01B

Instrument

Manual

PressCAL High Accuracy Digital

(Differential/Guage/Absolute)

Pressure Indicator

 
 

Summary of Contents for TDP-01B

Page 1: ...TDP 01B Instrument Manual PressCAL High Accuracy Digital Differential Guage Absolute Pressure Indicator...

Page 2: ...TABLE OF CONTENTS 1 Preface 3 2 Technical Specification 4 3 Understand the Instruments 5 3 Calibration 9 4 Warranty Certificate 14...

Page 3: ...2 Operating manual With Warranty certificate 3 Silicon pressure tubes 5 Carrying Bag STANDARED ACCESSORIES Welcome thanks for purchasing Our PressCAL Model No TDP 01B Our high Accuracy pressure measu...

Page 4: ...f 5 digits For higher ranges depends upon the pressure range Linearity Maximum 0 05 FS Defined as best fit straight line User selectable 12 multiple Pressure measurement units Kpa pa hPa Bar mBar Kg c...

Page 5: ...tting minimum maximum values Zero Key Enter Press for 1 second to zero tare the pressure display this key works as Enter command in program mode Light Key switch On Off for display light keep display...

Page 6: ...Battery compartment Remove Silicon protective cover to replace the battery Batteries AA size alkaline...

Page 7: ...m Autozero automatically for first time at the start a message will be displayed Autozero successfully done Users can also press zero key manually for 1 second to perform autozero manually User can ch...

Page 8: ...NG INTO SETTING PROGRAM MODE Enter password 1204 with help of shift left key Up arrow Increment key press Zero enter Key to enter the password If no key is pressed for long time the instrument will co...

Page 9: ...FS throughout the operating range Use this Mode only when you have calibrated external pressure reference source hi accuracy pressure display to calibrate this instrument Connect the External referenc...

Page 10: ...t XXXX Follow same steps at every 10 till step calibration is complete till SPAN a message will appear that calibration is complete ACTIVATION OF EDIT VALUES IN CALIBRATION MODE Edit Values correction...

Page 11: ...ion mode with help of enter key After the LCD Display shows the following prompt Edit Table 1 Pressure at Zero 2 Pressure at 10 3 Pressure at 20 4 Pressure at 30 5 Pressure at 40 6 Pressure at 50 7 Pr...

Page 12: ...left arrow Unit and up arrow key after editing of count value move to edit table with help of exit key light key After Show edit table in display inter Sr no 2 pressure at 10 with help of inter key Ze...

Page 13: ...key Zero key Repeat the same procedure for other values Edit the count value according to pressure value With help of left arrow Unit and up arrow key after editing of count value move to edit table w...

Page 14: ...hat Dry Block Temperature Calibrator having Sr No Date is propery Tested for workmanship We Certify our Calibrator for satisfactory performance for the period of one year from the date of supply again...