We recommend for meter cleaning and disinfection you should use
the disinfecting wipes / towelettes from below. The following product
with isopropyl alcohol as the active ingredient has been shown to be
safe for use with the Tula Health THG001 Blood Glucose Monitoring
System:
To obtain disinfecting wipes and other information, please contact
Medline at 1-800-MEDLINE (1-800-633-5463) or visit
www.medline.com. You can also purchase it at Amazon
(www.amazon.com) or Walmart (www.walmart.com).
Disinfecting Procedures
1. Wipe all meter’s exterior surface display and buttons by three
passes vertically and three passes horizontally with a folded
disinfecting towelette. Hold the meter with the test strip slot
pointing down and wipe the area around the test slot but be careful
not to allow excess liquid to get inside. Keep meter wet with
disinfection solution contained in the wipe for a minimum of 2
minutes for Micro-Kill+™ wipes.
2. Remove the wipe. Allow the meter surface to dry completely.
3. Discard the used wipes and never reuse them. Users should wash
hands thoroughly with soap and water after handling the meter,
lancing device, or test strips.
Improper system cleaning and disinfection may result in meter
malfunction. If you have a question, please contact local customer
service at
1-855-352-7884.
Micro-Kill+
™
(Micro-Kill Plus
™
) by Medline (EPA Reg. No. 59894-10-37549)