![TT Tooth Transformer TT Transformer User Manual Download Page 24](http://html.mh-extra.com/html/tt-tooth-transformer/tt-transformer/tt-transformer_user-manual_1178652024.webp)
Rev. 5 of 01.08.2020
24
13. TECHNICAL SPECIFICATIONS
TT TRANSFORMER is compliant with Council DirecAve 93/42/EEC of 14/06/93.
• External power supply voltage and frequency: 100-240 Vac, 50-60 Hz.
• Device power supply voltage max. absorbed power: 24 Vdc.
• Power: maximum absorbed power of 50 WaW.
• Processing cycle and warning messages.
OperaAng CondiAons
• OperaTng temperature: 5°C to 45°C (41° to 99° F).
• Storage temperature: –20°C to 60°C ( –4°F to 120°F).
• OperaTng humidity: max 80% up to 31°C (87°F), linear decrease to 50% 50% at
40°C (104°F).
• Storage humidity: <80%.
• OperaTng alTtude: 4000 meters (13,000 feet) maximum.
Weight and Dimensions
• Weight: 6,1 kg.
• Dimensions: 390mmx176mmx156 mm.
ElectromagneAc CompaAbility
:EN 61000-3-2, EN 61000-3-3, EN 61000-6-1, EN
61000-6-3.
14. ELECTROMAGNETIC COMPATIBILITY
EN 61000-3-2, EN 61000-3-3, EN 61000-6-1, EN 61000-6-3
WARNING:
contraindicaTons: interference from other equipment (an electro-scalpel or other
electrosurgical instruments in the proximity may interfere with the correct funcToning of the device).
WARNING:
portable and mobile radio-communicaTon sets may influence the correct funcToning of
the device.
1936
Summary of Contents for TT Transformer
Page 26: ...26 Rev 5 of 01 08 2020 ...