6. CLEANING OF THE MACHINE
Immediately aqer each use, TT Transformer must be cleaned.
WARNING
: you must always wear GLOVES and HALF-MASK during the cleaning procedure.
WARNING
: DO NOT remove the underlying drawer Waste when the machine is turned ON.
1.
Turn OFF the device.
2.
Extract from the machine the liquids tank and put it into its blister.
3.
Extract from the machine the Cylinder and put it into its blister.
WARNING
: manage carefully the Cylinder because now it is full of liquids.
4. Waste these components as described in paragraph “Disposal".
5. Extract the drawer named “Waste” from the machine and clean it.
6. Verify that in the Waste there is only dust.
7. Remove from the surface of the machine the transparent sTcker and clean the machine with a
soq gauze by saturaTng with a soluTon such as 75% isopropyl alcohol or other anTsepTc soluTon
cerTfied as medical device.
WARNING
: before next use, wait that surface are completely dry and there are not flammable
vapour of cleaning soluTons.
WARNING
: DO NOT submerge the device into liquid and DO NOT spray any liquid on it directly.
WARNING
: never use any sharp tool to remove dirty, since it will scratch the surface.
8. Clean the surface with a damp cloth and then dry with a piece of cloth.
9.
Sterilize Mill with autoclave, using universal program 121 °C / 2 bar / 16 min or,
alternaTvely, 134 °C / 2,1 bar / 5-7 min. The maximum recommended number of
sterilizaTon cycles in autoclave is 100 cycles.
Summary of Contents for TT Transformer
Page 26: ...26 Rev 5 of 01 08 2020 ...