(GATE EDGE MODE) This mode is available for hardware version 1~ (HD-VER 1~).
Count data is stored on the memory at the falling edge of GATE signal.
Counter starts at the first falling edge of GATE signal after STRT command.
START(COMM)
"H" time
"L" time
GATE(TTL IN)
COUNTING
COUNTING
COUNTING
LATCH DATA(INTERNAL)
Counter start
MEMORY SAVE(INTERNAL)
SAVE TIME
SAVE TIME
SAVE TIME
To meet the condition of SAVE TIME >= 10ms,
please input COUNTING TIME + SUSPENDING TIME >= 10ms.
4-2.Preparation for data acquisition commands
Data are stored on memory(Address:0~9999).
You can set up acquisition starting data address and end.
Data acquisition stops automatically if current data address reach the acquisition end data address.
You can read current data acquisition address.
4-2-1.Clear current data address
Type of command
CLGSDN
Clear current data address.
Set data acquisition starting address to zero.
4-2-2.Clear current data address and all stored data
Type of command
CLGSAL
Clear current data address and all memory data.
Set data acquisition start address to zero and clear all memory data.
4-2-3.Set and read gate synchronous data acquisition current data address
Type of command
GSDNddd・・・
Set current data address with a decimal value.
ddd・・・ : 0 ~ 9999
Type of command
GSDN?
Read out current data address.
Reply
0 ~ 9999