Using Serial Data Commands
C-25
Correlated (Paired) Data Record (C)
The C record is a multi-record report with a header (C0) followed by n more
records (C1...Cn)
CS,C,0,PNX,tindex,ffff,stime,dtime,evt1,evt3,evt4,total,n,m (the C records
only available in Correlated mode)
CS
= Checksum
C
= Correlated (Paired) Data Record
0
= 0 indicates header record for C data report
P
= C for Correlated Mode
N
= N for normal operation
A if in autocal mode
D if in au
tocal and autocal was “done” at the beginning of this
sample
X
= X for spare position
tindex
= time index 0 to sample time value - 1
(note if enabled for unpolled operation the
record is always reported when tindex = 0)
ffff
= 4 digit hex value for status flags (see RF command)
stime
= sample time not corrected for dead time
dtime
= dead time (ms for 3321, µs for 3320)
evt1
= number of single hump events
evt3
= number of 3+ hump events
evt4
= number of timer overflow events
total
= total (2-hump) particles measured (sum of reported channels,
evt 2)
n
= number of C (TOF) records to follow this record
m
= number of SS fields per record (set by SCR command)
CS,C,n,c1,c2,c3,... cm
CS
= Checksum
C
= correlated (paired) data record
n
= C1 to Cn records of correlated data where n is the number of
the aerodynamic dia. chan.
(each record contains data for one aerodynamic particle size
channel)
c1... cm
= particle counts in each pulse height accumulator bin (no
zeros)
Summary of Contents for SIZER 3321
Page 2: ......
Page 10: ...viii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 14: ...xii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 16: ...xiv Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 26: ...2 6 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 36: ...3 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 50: ...5 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 62: ...A 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 98: ......
Page 100: ......