C-16
Model 3321 Aerodynamic Particle Sizer
Spectrometer
R e a d C o m m a n d s ( P o l l e d )
Read commands are polled, which means the Model 3321 sends data in
response to a specific request from the computer
R
—Read Accumulator
This is a 3310 mode command. It dumps accumulator data taken using a G
command.
R
b,e
where:
b
= the beginning accumulator bin (0-1023).
e
= the ending accumulator bin (0-1023).
e
must be greater than or equal
to
b
.
If
b
and
e
are not specified, the default values are
b
= 0 and
e
= 1023. Line
feeds are appended after all carriage returns.
Example
Response
R0,100
Reads bins 0 to 100
R,100
Reads bins 0 to 100
R400
Reads bins 400 to 1023
R
Reads bins 0 to 1023
RF
—Read Flags
RF returns a four-character hexadecimal value representing the state of
the Model 3321.
The values for the 10 flags are as follows:
0000 0000 0000 0001
Laser fault
0000 0000 0000 0010
Total Flow out of range
0000 0000 0000 0100
Sheath Flow out of range
0000 0000 0000 1000
Excessive sample concentration (alarm)
0000 0000 0001 0000
Accumulator clipped (i.e. > 65535)
0000 0000 0010 0000
Autocal failed
0000 0000 0100 0000
Internal temperature < 10°C
0000 0000 1000 0000
Internal temperature > 40°C
0000 0001 0000 0000
Detector voltage more than ±10% Vb
0000 0010 0000 0000
Reserved (unused)
Summary of Contents for SIZER 3321
Page 2: ......
Page 10: ...viii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 14: ...xii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 16: ...xiv Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 26: ...2 6 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 36: ...3 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 50: ...5 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 62: ...A 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 98: ......
Page 100: ......