Using Serial Data Commands
C-21
UA
—Generate Accumulator Report Record
(Report Record A)
Enables and disables reporting of Record A in unpolled operation.
UAx
where:
x = 1 Enables A Record
x = 0 Disables A Record
UA will echo current setting 0 or 1.
See Record A in the “
” section of this chapter.
UB
—Generate Side Scatter Accumulator Report
Record (Report Record B)
Enables and disables reporting of Record B in unpolled operation.
UBx
where:
x = 1 Enables B Record
x = 0 Disables B Record
UB will echo current setting 0 or 1.
See Record B in the “
” section of this chapter.
UC
—Generate Correlated (Paired) Report Record
(Report Record C)
Enables and disables reporting of Record C in unpolled operation.
UCx
where:
x = 1 Enables C Record
x = 0 Disables C Record
UC will echo current setting 0 or 1.
See Record C in the “
” section of this chapter.
Summary of Contents for SIZER 3321
Page 2: ......
Page 10: ...viii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 14: ...xii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 16: ...xiv Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 26: ...2 6 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 36: ...3 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 50: ...5 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 62: ...A 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Page 98: ......
Page 100: ......