Sensor Alignment
RH Sensor Alignment
16
Sensor Alignment
For highly accurate measurement readings, you should align your QUESTemp
º
48N prior to data
logging or event logging/recording. For the RH sensor, it is recommended to align with a High
concentration level using the 75% sensor salt, NaCl (sodium chloride), or a Low concentration
level using the, the 33% sensor salt, MgCl (magnesium chloride). (To order sensor alignment
salts, refer to an online vendor such as, colepalmer.com.)
RH Sensor Alignment
1.
To open, navigate to the main menu and select
View
by pressing
I/O Enter
key (see
A
).
A measurement screen will appear (see
B
).
A
B
VIEW
SETUP
RESET
WET
50.7° F
DRY
70.6° F
Figure 1-24: Main menu with view selected (A) & measurement screen (B)
2.
Press and hold
I/O Enter
key and then press
Down Arrow
key from the
View
menu. The
Alignment
screen will appear.
(RH Sensor) Alignment menu
Alignment RH
WBGT
HIGH LOW
“High” is selected when the arrow appears in
front of the text.
NOTE
: WBGT is not used in sensor alignment
Figure 1-25: Alignment screen
3.
Press
Up/Down
Arrow
key to select either
High
or
Low
. Then press
I/O Enter
key.
4.
Remove
(or slide) the
Sunshield
(white globe) from the
RH sensor
and place it to the side.
(See Figure 1-26).
Sensor
Alignment