Background Info:

The DRX can be powered by battery (up to two battery packs inserted in the rear of the unit) or by DC voltage supplied by a
battery/solar panel system or inverter from a 120V or 240V source through the power connector on the right side of the unit
when it is facing towards you. The unit has internal memory for around 90 days at 2-minute averaging intervals (the minimum
required recording interval to utilize the auto-zero unit), which can be downloaded directly from the DRX instrument with a USB
drive OR with a PC using the TSI TrakPro software via supplied USB cable connected to the USB Device connection on the device.
Note: the only option to download using a USB drive is download all and only in a comma separated file format.
The DRX device communicates through NDIS IP-based drivers and therefore marries to a PC device if kept ON. This means if you
go to download from a DRX device with a different PC it will not work without stopping and restarting the DRX device.

Environmental Enclosure:

The DRX device will reside in the environmental enclosure to protect it from the elements. It is strapped into the enclosure using
Velcro, attached to the inlet assembly with tubing, connected to the power source through the power cable on the right side of
the enclosure, and a USB connector to the device connector on the DRX device is accessible via a port on the left side of the
enclosure.

Auto-Zero Unit

Summary of Contents for DustTrak DRX

Page 1: ...k Export Name file based on the date as YYYYMMDD 17 Now delete the test files by going to File Receive select Delete All 18 Correct the time of the DRX through Setup Parameters Clock 19 Note the time...

Page 2: ...nnected to the USB Device connection on the device Note the only option to download using a USB drive is download all and only in a comma separated file format The DRX device communicates through NDIS...

Page 3: ...n Settings are more easily modified through a PC using the TrakPro software which will be covered in the next section 5 Go into the Zero Cal option be selecting it This is where one can manually zero...

Page 4: ...of the connected DRX device c If the above happens go to back into the Logging Control following step 3 d If the above does not happen wait a couple seconds and click the Test button e If the option i...

Page 5: ...a 2 minute recording interval choose a Start Time of 14 10 or 14 12 or 14 14 or 14 16 etc c Select Send on the right hand side once the options are correct 13 To actually start the DRX device you stil...

Page 6: ...fore it starts to record iii Log Interval This is how often the data is recorded in the final memory For a 2 minute data interval required this should be set to 2 minutes iv Test Length This is how lo...

Page 7: ...sten to the pump while running the DRX device in Manual RunMode Under Setup FlowCal on the DRX unit interface the speed of the pump can be increased decreased make sure there is an audible coordinatin...

Reviews: