TSI Instruments DUSTTRAK 854001-M1 Assembly Manual Download Page 9

–9– 

14. If using a Thiamis 1000 Node, route the other side of 

the cable to the power input on the Thiamis node 
(green connector). 

 

Heated Inlet Operating Instructions  

1.  When powered and attached to the DustTrak

™ monitor, the Heated Inlet will automatically function. 

2.  Select RH set point that will be controlled at the entrance to DustTrak

™ monitor to be 30, 40 or 50% 

RH via the DIP switches. Heated inlet will then power the heater to heat incoming air and thereby 
decrease RH to the targeted level. 

NOTE 

By default, TSI

®

 Incorporated (TSI

®

) sets the RH set point to 30% RH at the factory. If needed, this 

can be adjusted to meet other application needs/requirements. 

 

 

 

The 

LED 

describes the status of the heated inlet: 

Solid Green 

Inlet temp. is < 1°C below set point and controlling to maintain set point. 

Blinking Green 

Inlet temp. is between 1 to 5°C below set point and controlling to improve. 

Blinking Red 

Inlet temp. is more than 5°C below set point and controlling to improve. This will 
occur when unit is first turned on and coming to temperature. 

Solid Red 

Sensor unplugged or has issue. 

Summary of Contents for DUSTTRAK 854001-M1

Page 1: ...DUSTTRAK ENVIRONMENTAL MONITOR MCERTS VERSIONS ASSEMBLY GUIDE P N 6010298 REVISION C FEBRUARY 2021...

Page 2: ...mes and mist This instrument use a sheath air system that isolates the aerosol in the optics chamber to keep the optics clean for improved reliability and low maintenance The below models are designed...

Page 3: ...instrument is run with these Photometric Correction Factors However the application of site specific correction factors may be used if a significant decrease in the expanded uncertainty can be demons...

Page 4: ...Omni Directional Inlet as shown NOTE For best results use PTFE Thread Sealant Tape on the water bottle threads 2 Unscrew the PM2 5 or PM10 impactor to access the impactor plate as it comes pre assembl...

Page 5: ...Sample Conditioner column by pushing it down 6 Ensure the O ring included with the Impactor Kit is installed into the bottom of the Impactor housing 7 Thread hand tighten the impactor housing with O...

Page 6: ...Heated Inlet 854041 into Environmental Enclosure 1 Attach heated inlet control module to the front of the Photometer This can be done using the 2x 6 32 x 5 16 Socket Head Cap Screws and 2x flat washe...

Page 7: ...wn Sun Shield should not touch sintered cap of RH Temp Probe 5 Attach RH Temp Sensor cable to heated inlet control module J3 6 Attach the two larger gray connectors to side of the photometer 7 Attach...

Page 8: ...nting ring rotate slowly until alignment pin mates with the hole in the bottom of heated inlet column 10 Slide column retainer ring down heated inlet column 11 Hand tighten column retainer ring to sec...

Page 9: ...oming air and thereby decrease RH to the targeted level NOTE By default TSI Incorporated TSI sets the RH set point to 30 RH at the factory If needed this can be adjusted to meet other application need...

Page 10: ...trol module 7 in x 1 5 in x 0 75 in 17 8 cm x 3 8 cm x 1 9 cm Temp Range 0 to 50 C 32 to 122 F RH Range 0 to 95 RH Set Points 30 RH 40 RH 50 RH Warm up Time 17 minutes CE IEC 61326 IEC 61010 1 Instrum...

Page 11: ...11 This page intentionally left blank...

Page 12: ...strations DustTrak is a trademark of TSI Incorporated TSI Incorporated Visit our website www tsi com for more information USA Tel 1 800 680 1220 UK Tel 44 149 4 459200 France Tel 33 1 41 19 21 99 Germ...

Reviews: