–8– 

8.  Install O-ring and route cable through the mounting 

ring. Pull excess cable through to prevent pinching 
of cable when inlet column is installed.  

 

9.  Set inlet column on top of 

photometer mounting ring, rotate 
slowly until alignment pin mates 
with the hole in the bottom of 
heated inlet column. 

10.  Slide column retainer ring down 

heated inlet column. 

11.  Hand-tighten column retainer ring 

to secure inlet column to 
photometer mounting ring.  

  

 

12. Attach ferrite to cable with one wrap of the cable and 

snap the ferrite together. 

 

13. Attach heated inlet cable to heated inlet control 

module (J4).  

 

Summary of Contents for DUSTTRAK 854001-M1

Page 1: ...DUSTTRAK ENVIRONMENTAL MONITOR MCERTS VERSIONS ASSEMBLY GUIDE P N 6010298 REVISION C FEBRUARY 2021...

Page 2: ...mes and mist This instrument use a sheath air system that isolates the aerosol in the optics chamber to keep the optics clean for improved reliability and low maintenance The below models are designed...

Page 3: ...instrument is run with these Photometric Correction Factors However the application of site specific correction factors may be used if a significant decrease in the expanded uncertainty can be demons...

Page 4: ...Omni Directional Inlet as shown NOTE For best results use PTFE Thread Sealant Tape on the water bottle threads 2 Unscrew the PM2 5 or PM10 impactor to access the impactor plate as it comes pre assembl...

Page 5: ...Sample Conditioner column by pushing it down 6 Ensure the O ring included with the Impactor Kit is installed into the bottom of the Impactor housing 7 Thread hand tighten the impactor housing with O...

Page 6: ...Heated Inlet 854041 into Environmental Enclosure 1 Attach heated inlet control module to the front of the Photometer This can be done using the 2x 6 32 x 5 16 Socket Head Cap Screws and 2x flat washe...

Page 7: ...wn Sun Shield should not touch sintered cap of RH Temp Probe 5 Attach RH Temp Sensor cable to heated inlet control module J3 6 Attach the two larger gray connectors to side of the photometer 7 Attach...

Page 8: ...nting ring rotate slowly until alignment pin mates with the hole in the bottom of heated inlet column 10 Slide column retainer ring down heated inlet column 11 Hand tighten column retainer ring to sec...

Page 9: ...oming air and thereby decrease RH to the targeted level NOTE By default TSI Incorporated TSI sets the RH set point to 30 RH at the factory If needed this can be adjusted to meet other application need...

Page 10: ...trol module 7 in x 1 5 in x 0 75 in 17 8 cm x 3 8 cm x 1 9 cm Temp Range 0 to 50 C 32 to 122 F RH Range 0 to 95 RH Set Points 30 RH 40 RH 50 RH Warm up Time 17 minutes CE IEC 61326 IEC 61010 1 Instrum...

Page 11: ...11 This page intentionally left blank...

Page 12: ...strations DustTrak is a trademark of TSI Incorporated TSI Incorporated Visit our website www tsi com for more information USA Tel 1 800 680 1220 UK Tel 44 149 4 459200 France Tel 33 1 41 19 21 99 Germ...

Reviews: