Appendix B
62
Downloading a Data File
To download a file under normal data-logging operations, you would do the
following:
•
Open the Environmental Enclosure.
•
Discontinue sampling.
•
Attach the supplied RJ45 to the RS-232 cable from the D
UST
T
RAK
monitor to a notebook computer.
•
Download the data file using T
RAK
P
RO
this manual for further details).
Real-Time Monitoring
The Environmental Enclosure comes equipped with an external DB9
connector. This allows the D
UST
T
RAK
monitor to be connected to a
computer for continuous real-time monitoring. To set up the Environmental
Enclosure for real-time monitoring follow these instructions:
•
Connect the D
UST
T
RAK
monitor to the DB9 connector, which is
mounted on the inside of the enclosure, using the Computer Interface
Cable provided with the D
UST
T
RAK
monitor
.
•
Attach a standard 9-pin serial cable (not supplied) from the DB9
connector on the outside of the enclosure to a computer serial port. The
computer can be configured to query the D
UST
T
RAK
monitor when in a
terminal emulator mode. See Application Note ITI-044 for more
information on these commands.
Setting-up T
RAK
P
RO
TM
Data Analysis Software
Information on setting-up and using T
RAK
P
RO
software can be found in
.
Summary of Contents for DustTrak 8520