Chapter 3
36
2.
Select
DustTrak Model 8520
from the
Options
menu under
Software
Configure
. Or, select
DustTrak
from list of available instruments on
the drop-down menu on the menu bar.
3.
Open the file that contains the data to be converted. Highlight a single
test within that data file.
4.
Select
Define DustTrak Calibration
from the
Options
menu. The
following dialog box is displayed.
5.
Enter the
New Mass Concentration
value (in mg/m
3
). The
New
Calibration Factor
is calculated and automatically updated. Press
OK
.
6.
When you click OK, a new data set of the
same name
will be generated
calibrated to the mass concentration that you entered. This data set will
be appended to the same file. The original data remains.
7.
If you want to adjust the data to a previously determined custom
calibration factor without performing a gravimetric calibration, enter the
new calibration factor. The
New Mass Concentration
automatically
updates.
05/09/2006,
Summary of Contents for DustTrak 8520
Page 2: ......
Page 10: ...vi ...
Page 14: ...Chapter 1 4 ...
Page 20: ...Chapter 2 10 ...
Page 82: ......
Page 84: ......
Page 88: ...Index 78 W X Y warning v warranty 2 wrist strap 1 3 Z zero checking 38 zero filter 1 ...