36
Chapter 3
Frame Kit
1081263
It is recommended that you purchase a frame kit whenever you
purchase a hood other than the 2 ft x 2 ft size. The frame kit contains
all extra frame tubing and connectors required to build frames for the
various size hoods. The frame kit consists of the following parts:
4 pieces 1-foot frame tubing
2 pieces 2-foot frame tubing
6 pieces straight tubing connectors
Meter
The A
CCU
B
ALANCE
has a built-in electronic flow meter. Do not try to
detach the meter from the base. The A
CCU
B
ALANCE
should be used
and stored within the specified temperature range, 32 - 140º F (0-60º
C). The meter case, display screen and membrane switch can be
cleaned using a damp cloth with mild detergent solution. Do not
immerse the meter in water. Wipe the meter dry before use.
Manifold
If you notice that a tube in the flow sensing manifold has become
disengaged from its rubber holder as shown in Figure 14, work the
tube end back into the holder. The accuracy of the A
CCU
B
ALANCE
is
not affected by having one or two tubes out of their holders; however,
it is a good practice to keep the tubes properly in place.