Firmware Commands
B-17
Example D Record:
D,2,0,2.27e3,6.0,5.875,66784,0,308
S Record (status)
S, FL, AP,GT,OT,CT
S
Status Record Identifier
FL
Aerosol flow rate in cm
3
/min
AP
Absolute Pressure in millibar
ST
Saturator Temperature in degrees Celsius
GT
Growth tube Temperature in degrees Celsius
OT
Optics Temperature in degrees Celsius
Example S Record:
S,300,970,12.0,75.0,75.0
Z Record (scanning data)
Z,Mode,Flags,RN,N0,N1… ,N9
Z Record
Identifier
Mode Current Operating mode (5 or 6)
Flags See status flags
RN
Record number - (1 to 600)
Nx
Equivalent number of particles counted in tenth second (0 to 99999999) based on live-
corrected counts or the photometric value.
Example Z record:
Z,5,0,12,0,0,0,12,182,518,641,896,887,871
A Record (raw analog -- diagnostic only)
A,a,b,c,d,e,f,g,h,i,j,k,l,m,n,o
A Record
Identifier
a-o raw analog values (bits)
Example A record:
A,2382,3194,1,1108,4095,4095,2735,1284,518,641,896,2278,821,819,1958
C Record (flow and temperature control - diagnostic only)
C,a,b,c,d,e,f,g,h,i,j,k,l
C Record
Identifier
a - l control bit values
Example C record:
C,2413,2422,2166,2276,2276,1468,821,820,2894,821,820,1011,2210,2222,2012