Firmware Commands
B-9
SIT
S
et
I
njection
T
ime
This command sets the water injection rate in seconds between injections. The factory
default setting is 12 seconds.
Command:
SIT,xxxx
Set Injection Time command
xxxx ranges 0 to 3600 seconds
Response:
OK
Example:
SIT,12
Command sets injection interval to 12 seconds
OK
Response
SIT
Command to read back the current setting
12
Response
SL
S
et
L
aser
The set laser command switches the laser on or off. The laser is always in the “on”
condition when the instrument is first powered up.
Command:
SL,x
Command to set laser on or off
x = 0 turns laser off, x = 1 turns laser on
Response:
OK
Example:
SL,0
Command turns laser off
OK
Response
SL
Command to read back the current setting
0
Response indicates laser is off
SLM
S
et
L
ive-time
M
inimum
This command sets minimum amount of live-time in proportion to elapse time allowed
before the “OVER” annotation is shown on the primary display. The factory default
value is 400 which represents 40%.
Command:
SLM,xxx
Command to set live-time minimum
xxx = 0 to 1001
Response:
OK
Response
Example:
SLM,400
Command to set live-time minimum to 40%
OK
Response
SLM
Command to read back the current setting
400
Response