60 Chapter
7
Symptom
Possible Causes
Solution
Fails Zero Check
(continued)
O-ring on alcohol
cartridge is not sealing.
Replace O-ring. Smear a
very
small amount of
grease (such as
Vaseline™) on the O-
ring.
Tube fittings on
P
ORTA
C
OUNT
®
P
RO
fit
tester are loose.
Tighten fittings with
pliers.
Fit factor of 1 or
very low
(If the
P
ORTA
C
OUNT
passes the Daily
Checks, the
problem is with
the respirator, not
the
P
ORTA
C
OUNT
.)
Respirator is not
equipped with HEPA,
class-99, class-100, or
P3 filters.
Respirator leaks, has
loose filters, or a
malfunctioning
exhalation valve.
Twin Tube Assembly is
disconnected.
Install proper filter for fit
testing.
Repair the respirator.
Connect the Twin Tube
Assembly to the
P
ORTA
C
OUNT
®
P
RO
fit
tester.
Twin Tube Assembly is
not connected to
respirator sampling port.
Connect the Twin Tube
Assembly to the respirator
sampling port (clear
tube).
Twin Tube Assembly
Leaks.
Repair or replace the
Twin Tube Assembly.
Sample Tube too Long.
Use standard 5-foot Twin
Tube Assembly.
P
ORTA
C
OUNT
®
P
RO
fit
tester is flooded with
alcohol.
Run 2 hours to dry optics
(remove the Alcohol
Cartridge and install
Storage Cap).
Alcohol level is low.
Add alcohol to the
P
ORTA
C
OUNT
®
P
RO
.
Switching valve is not
functioning.
Return to TSI for service.