28 Chapter
4
9.
Set the Alcohol Fill Capsule down and wait at least 2 minutes while the
alcohol wick inside the Alcohol Cartridge soaks up alcohol.
10.
Remove the Alcohol Cartridge from the Capsule and gently shake it to
allow excess alcohol to drip back into the Alcohol Fill Capsule. Stop
when excess alcohol is no longer dripping; it is not necessary to wait
until the outside surface of the Alcohol Cartridge is dry.
11.
Insert the Alcohol Cartridge into the Cartridge Cavity of the
PORTACOUNT® Pro Respirator Fit Tester. It should slide in with little
effort. DO NOT FORCE IT. As you approach full insertion, firmly twist
the Alcohol Cartridge clockwise until it locks into position.
12.
Recap the alcohol fill capsule with the storage cap.