38
Certifier
®
FA Plus Test System
Symptom
Possible Cause
Corrective Action
Volume, minute
volume, peak flow,
peak pressure, PEEP,
respiratory rate, or I:E
ratio measurement are
not updated or
incorrect.
Less than two
consecutive full breaths
have been supplied to
flow module, or flow is not
supplied as a breathing
waveform.
Auto-trigger not optimized
for current waveform.
Airway restrictor causing
flow disturbance next to
flow module.
Flow condition causing
oscillations in the
waveform.
Wait for at least two consecutive
full breaths to be supplied to the
flow module.
Ensure that flow is supplied as a
breathing waveform.
Manually set flow trigger level.
Use the TSI Airway pressure
fitting which contains the screen
between the flow module and
the test lung (see Figure 12).
Minimize length of tubing
between the test lung and the
airway pressure fitting.
I:E ratio or I time
readings do not
appear correct.
It is possible the
inspiratory pause function
is turned on inside the
ventilator.
Select the I time parameter T
I+P
and ratio parameter I:E
IP
to be
on the display when the
inspiratory pause feature is
activated on the ventilator. See
section 2.3.3 for a description of
the difference between T
I
and
T
I+p
.
Another option is to turn off the
inspiratory pause feature on the
ventilator.
Can’t zero low-
pressure or high
pressure transducer.
Transducer is connected
to a pressure source.
Disconnect pressure tubing from
flow module and then zero low-
pressure transducer.
Can’t zero barometric
pressure transducer.
Barometric pressure
transducer does not
require a zero calibration.
Resume normal system
operation.
Summary of Contents for Certifier FA Plus
Page 1: ...CERTIFIER FA PLUS TEST SYSTEM OPERATOR S MANUAL P N 1980560 REVISION J FEBRUARY 2016...
Page 2: ......
Page 4: ......
Page 10: ...vi Certifier FA Plus Test System This page intentionally left blank...
Page 17: ...1 Introduction 7 Figure 5 Interface Module Touch Screen Display Key Pad Stylus...
Page 20: ...Certifier FA Plus Test System 10 Figure 9 Oxygen Sensor Kit...
Page 50: ...40 Certifier FA Plus Test System This page intentionally left blank...
Page 58: ...48 Certifier FA Plus Test System This page intentionally left blank...