24
Certifier
®
FA Plus Test System
2.3.4
Gas Conditions Selection Box
Figure 22. Gas Conditions Selection Box
STP
Standard Temperature and Pressure
. The gas flow rate and volumes
are displayed in terms of what the gas flow rate and volume would be if
the gas was 21°C and 1 atmosphere (101.3 kPa) of pressure.
ATP
Actual Temperature and Pressure
. The gas flow rate and volumes at
the actual temperature and pressure of the gas.
BTPS
Body temperature and Pressure Saturated
. The gas flow rate and
volumes are displayed in terms of what the gas flow rate and volume
would be if the gas was changed to 37°C, the actual pressure, and also
saturated with water vapor.
BTPD
Body temperature and Pressure Dry
The gas flow rate and volumes
are displayed in terms of what the gas flow rate and volume would be if
the gas was changed to 37°C, but maintaining the actual pressure.
STPxx
User defined Standard Conditions
. The gas flow rate and volumes are
displayed in terms of what the gas flow rate and volume would be if the
gas was at the conditions entered in the boxes to the right.
Summary of Contents for Certifier FA Plus
Page 1: ...CERTIFIER FA PLUS TEST SYSTEM OPERATOR S MANUAL P N 1980560 REVISION J FEBRUARY 2016...
Page 2: ......
Page 4: ......
Page 10: ...vi Certifier FA Plus Test System This page intentionally left blank...
Page 17: ...1 Introduction 7 Figure 5 Interface Module Touch Screen Display Key Pad Stylus...
Page 20: ...Certifier FA Plus Test System 10 Figure 9 Oxygen Sensor Kit...
Page 50: ...40 Certifier FA Plus Test System This page intentionally left blank...
Page 58: ...48 Certifier FA Plus Test System This page intentionally left blank...