Instrument Operation
4-7
P e r f o r m i n g a F i l t e r T e s t
Performing a filter test is very similar to performing fit testing described in
the previous section. The UPSTREAM sample is drawn from the filter
holder, providing the challenge aerosol. The DOWNSTREAM sample is
taken after the filter.
Filter efficiency is calculated as a percent from the following equation;
n
Penetratio
%
100
Efficiency
%
voltage
Zero
voltage
Upstream
voltage
Zero
voltage
Downstream
n
Penetratio
100
voltage
Zero
-
voltage
Upstream
voltage
Zero
-
voltage
Downstream
n
Penetratio
%