![TSI Incorporated 8026 Operation And Service Manual Download Page 14](http://html1.mh-extra.com/html/tsi-incorporated/8026/8026_operation-and-service-manual_1178063014.webp)
Chapter 3
8
2
3
4
5
1
Figure 3-1
3. Use mild soap and warm water to clean the jet assembly parts.
Examine the orifice for plugging. Rinse a second time if necessary.
Either use clean compressed air to dry the parts, or air dry them.
Note
: The orifice is easily damaged. Do
not
insert anything in the
orifice in an attempt to clean it.
4. Reassemble and install the atomizer jet, making sure to replace
the O-ring. When replacing the retainer, make sure it is pushed in
completely.
5. Attach the draw tube.
6. Test the generator for proper operation (output) using clean tap
water.