![TSI Incorporated 8026 Operation And Service Manual Download Page 13](http://html1.mh-extra.com/html/tsi-incorporated/8026/8026_operation-and-service-manual_1178063013.webp)
7
Chapter 3
Maintenance
Routine Maintenance
The Model 8026 Particle Generator requires very little maintenance.
Cleanliness is the single most important maintenance requirement.
To ensure trouble-free operation of the generator, do the following:
Use compressed air to blow out accumulated salt from inside the
generator case. Blow air into the louvers located on the back of
the generator to clean salt from the cooling fan. Blow air into the
bottom of the generator through the cooling louvers.
C a u t i o n
This cleaning maintenance step is critical to maintaining
trouble-free operation of the generator.
Wipe up any salt solution spills immediately.
Periodically wipe down the outside surface of the generator using
a damp cloth.
After each use, shut off the generator, remove the salt solution
reservoir, and then switch the generator back on for 40 seconds.
This will purge salt solution from the unit and reduce
contamination of the carrying case.
No other routine periodic maintenance is required. If a decrease in
output is noticed, clean the atomizer nozzle by following the
procedure below.
Cleaning the Atomizer Nozzle
1. Turn the particle generator off. Remove the salt solution
reservoir.
2. Disassemble the atomizer jet assembly following the order
indicated in Figure 3-1. The retainer that holds the parts of the jet
together can be difficult to remove. If necessary, use pliers to pull
down on the tab with a
gentle
rocking motion. The jet assembly
incorporates one small O-ring. Set it in a safe place.