Specifications
15
Notes:
1
Accuracy stated at standard conditions of 21.1°C and 101.3 kPa.
Add an additional 0.0003 Std L/min or 0.075% of reading, whichever
is greater, per 1°C away from standard conditions when operating
within the range of 0°C to 50°C.
Add an additional 0.015% of reading per 1 kPa above 101.3 kPa or
Add an additional 0.022% of reading per 1 kPa below 101.3 kPa
when operating within the pressure range of 70 kPa to 101.3 kPa.
2
Accuracy stated with gas temperature and flow body temperature within
±10°C of one another.
3
Accuracy stated measuring dry gas (less than 10% R.H.).
4
Includes ±0.5% of reading repeatability.
5
Volumetric flow rate is calculated from the mass flow measurement. Add
an additional 0.25% of reading to the flow accuracy to account for the
uncertainty in measuring gas temperature and pressure.
6
For analog output, accuracy offset increases from 0.005 to .02 Std L/min.
7
Add uncertainty of 0.2 kPa for every 10°C away from 21.1°C.
*Specifications subject to change without notice.
(continued on next page)