background image

Intended to use:
The control solution is used to check the performance of the TRULY 
system, including Blood Glucose and Cholesterol Monitor System, 
Glucose and Cholesterol strip, user's test skill. The system is 
performing adequately if the control solution test result falls within the 
indicated control range listed on the test strips vial label.

A control test should be proceeded:
1. When you suspect that the meter or test strips are not working 
    properly.
2. When your blood glucose or cholesterol test results are not 
    consistent with how you feel, or when you think your results are not 
    accurate.
3. When you drop or bump your meter.
4. Upon the advice of the healthcare professional.

How to perform a control test.
1. Insert a test strip into the test strip port of the meter and immediately 
    cover the vial to keep the strips in dry condition.
2. Be careful to make sure the test mode and code number on the 
    screen are the same as the one marked on the test strip vial. If not, 
    remove the test strip and re-code the meter to correct test mode and 
    code number.
3. Shake the control Solution vial vigorously then open the cap to 
    discard the first three drops of control solution, after wiping the 
    dispense tip and cleaning with tissue then drop some control solution 
    onto a clean object such as transparency. If it is a cholesterol test, 
    Please drop out  enough control solution (at least 10uL).
4. Touch the Sample Inlet of the Glucose or Cholesterol Test Strip to the 
    drop of Control Solution until the Meter sounds a “beep”, 
    The screen will display a count down from 10 seconds for Glucose 
    Test Strip or 26 seconds for Cholesterol Test Strip.    
5. After counting down, the Screen will display the control Test results, 
    Compare the results to the control range listed on the Test Strip Vial.

Warning:

-13-

Chapter 4 Quality Control

Summary of Contents for TBG-460

Page 1: ...Blood Glucose Cholesterol 2 in 1 Monitoring System User s Manual 2015 07 14 Rev 01 www trulyinstrument com Model TBG 460GC...

Page 2: ...Testing 3 2 Connect iOS device via Bluetooth 3 3 Perform the Blood Glucose Test 3 5 Special messages 3 6 Care after the test Chapter 4 Chapter 5 Meter Memory Function Chapter 6 Installing Replacing th...

Page 3: ...ugh the operation procedures with illustrated directions and help you find the required informations Please contact the local distributor for questions and enquiries WARNING Please carefully read the...

Page 4: ...t Hct above 55 may cause reduced glucose results versus a whole blood reference Do not drop the product Keep the lancets and lancing device out of the reach of the children o Keep the meter and test s...

Page 5: ...companying documents Storage temperature limitation Use by EU Representative Catalogue number Product number Manufactured by 0123 1 5 System Components Test Strip Port Insert test strip here The meter...

Page 6: ...At least 10uL of blood sample is essential Reaction Zone Reaction zone of Total Cholesterol CHOL Glucose Test Strip Blue Color Sample Inlet The inlet where the whole blood or control solution is drawn...

Page 7: ...s are listed as follows Set the Year 1 When the meter enter setting mode the first item on the screen such as 2012 will blink on screen which means Year 2 Press the up down button to 3 After you set t...

Page 8: ...wm button until the correct measurement unit appears 2 Press the memory button will show OFF on the screen and power off Chapter 3 How to perform Test 3 1 Before Testing 7 Coding the Meter TRULY TRULY...

Page 9: ...so the same as marked on the test vial Remove the code strip should appear on the screen indicating that the meter has been successfully coded Cautions If E E appears on the screen during coding proce...

Page 10: ...napping it back onto the lancing device body 4 Select the skin penetration depth by turning the depth selector until the desired number can be seen in the selection window 5 levels of skin penetration...

Page 11: ...Be careful to make sure the marked re code test mode and code number on the screen are the same as the one on the test strip vial Icon GLUC and CHOL indicate measuring mode of glucose and total chole...

Page 12: ...wn from 26 seconds for cholesterol testing 3 4 Perform the Total Cholesterol Test 11 4 Touch the sample Inlet of the strip to the drop of blood sample until the meter sounds a beep which means the sam...

Page 13: ...s 3 6 Care after the Test Directions 1 Improper usage of the meter such as dropping bumping or other violent impact may cause the meter damaged 2 Please keep meter away by interferences of magnetic el...

Page 14: ...o keep the strips in dry condition 2 Be careful to make sure the test mode and code number on the screen are the same as the one marked on the test strip vial If not remove the test strip and re code...

Page 15: ...est mode only total cholesterol test mode do not have the day average function Chapter 5 Meter Memory Function Please read this user s manual for control solution test carefully before use Do not reus...

Page 16: ...l hear a beep sound which indicates that meter is 3 Place back the cover onto the meter ready 5 After the 28day average the latest stored result M01 appears on the screen Then you can read the stored...

Page 17: ...bent hole Revlew the instructiona and try again with a new test strip If the problem persists please contact the dealer for service E 9 Chapter 7 Storage and Handling of Meter and Strips Due to sensit...

Page 18: ...ter and the test strip have reached at temperature within the operating range E U It could be caused by a used test strip or a temporary or permanent electronic problem Repeat the test with a new test...

Page 19: ...se level 75 mg dL 4 17mmol L 20 at Total Cholesterol 150mg dL 3 88mmol L Electrochemical Biosensor Plasma equivalent Glucose 20 600 mg dL 1 1 33 3 mmol L Total Cholesterol 100 400 mg dL 2 59 10 35 mmo...

Page 20: ...radio frequency energy However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or reception which can be...

Page 21: ...Knitting Centre 1 3 Wing Yip St Kwai Chung N T HONG KONG Tel 852 2487 9803 Fax 852 2146 5756 Http www trulyinstrument com EC Representative Wellkang Ltd t a Wellkang Tech Consulting Suite B 29 Harley...