OXXIOM RX APP TECHNICAL
SPECIFICATIONS, CONT.
Delay Description
Max. Delay (seconds)
Data processing
12
Warning detection and
generation
30 (i.e., 12 + 18)
Measurement update
1
Notes
(1)
The Oxxiom Pulse Oximetry System applies True
Wearables’ proprietary algorithms to real-time
signals in order to calculate SpO2, PR and PA
measurements. These algorithms produce an
effective data processing time delay that is less
than or equal to 12 seconds. Motion in the
measurement site, and/or interferences, and/or low
perfusion, and/or rapid non-physiological in nature
changes in SpO2, PR, and/or PA levels may cause
additional delays of up to 18 seconds, which may
increase warning detection and generation delays.
Measurement gauges and trends are updated once
a second. The following summarizes the Oxxiom
Pulse Oximetry System absolute maximum delays: