OXXIOM DEVICE TECHNICAL
SPECIFICATIONS, CONT.
Notes
(4)
± 3 bpm is the root-mean-square accuracy, and
represents approximately 68% of measurements.
Evaluated with electronic pulse simulator study,
and during induced hypoxia study with human
subjects. The studies followed the pulse oximetry
guidelines of ISO 80601-2-61:2011 applicable
sections. Please refer to DEVICE
PERFORMANCE section in this guide for a
detailed description regarding Oxxiom’s PR
accuracy evaluation.
(5)
The Oxxiom Pulse Oximetry System has been
certified to comply with the following safety
standards:
•
ANSI/AAMI ES 60601-1:2005/(R) 2012: General
requirements for basic safety and essential
performance
•
IEC 60601-1-2:2014: General requirements for
basic safety and essential performance -
Collateral Standard: Electromagnetic
disturbances - Requirements and tests.