Model 3440
Appendix A
–
9
the alkaline batteries
need replacing.
Moisture Offset
Exceeds Limits!
The moisture offset
(k) value must be
between
–
200 and
200. When
performing a
moisture offset, the
resulting k value was
outside of these
limits.
Enter a new moisture offset.
No active project! Operator tried to
store measurement
data, but no project is
active.
Select an existing project or
create a new project and select
as active.
Proctor Target
Exceeds Limits!
The target value must
be between 20.0 and
200.0 pcf. The
operator has entered
a value outside this
range.
Enter a target value between
20.0 and 200.0 pcf.
Project has No
Data!
Operator attempted
to view a project with
no data.
Select the correct project when
attempting to view.
Overlay Thickness
Exceeds Limits!
When using the
Nomograph
function,
the overlay thickness
value must be
between 0 and 10 in.
The operator has
entered a value
outside this range.
Enter an overlay thickness
value between 0 and 10 in.
Warning!! Time
to Reach
Precision too
high
When using the
Precision
function, the
time required to reach
the selected precision
value
exceeds 60
minutes.
Enter a different precision
value or disable the Precision
function.
Summary of Contents for RoadReader 3430
Page 14: ...1 6 Figure 1 Model 3440 Gauge and Accessories...
Page 19: ...Model 3440 2 3 Figure 2 Direct Transmission Geometry Figure 3 Backscatter Geometry...
Page 20: ...2 4 Figure 4 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 22: ...2 6 Figure 5 Effect of Moisture on Depth of Measurement...
Page 164: ...F 2 STANDARD COUNT LOG Gauge Serial Number _____________ Date MS DS Date MS DS...