![Troxler Enhanced RoadReader Plus 3451 Manual Of Operation And Instruction Download Page 90](http://html1.mh-extra.com/html/troxler/enhanced-roadreader-plus-3451/enhanced-roadreader-plus-3451_manual-of-operation-and-instruction_1172542090.webp)
Model 3451
5–5
MODE
The Model 3451 provides three measurement modes (
Soil
,
Asphalt
,
and
Thin Layer
) for precise readings on base and thin layer asphalt,
concrete, soil, soil-stone aggregate, and similar materials.
Use the
Soil
mode when measuring soil or a soil-stone
aggregate.
Use
Asphalt
mode when measuring concrete or asphalt layers of
4 inches or greater.
Select the
Thin Layer
mode when measuring thin-layer asphalt
of less than 4 inches. In thin layer mode, the gauge can average
up to 20 readings from a single test location. When using this
function, the PDA automatically prompts for repositioning of
the gauge from a test location and begins another measurement.
After all measurements have been completed, the gauge
calculates the average density of the test material. When taking
counts on thin layer asphalt, Troxler recommends averaging
four one-minute readings.
NOTE
The Model 3451 gauge can store measurements
performed using different modes under a single project
name. For example, the operator may select the Soil
mode when setting up the project (see page 3–19), but
can also save Asphalt mode measurements under the
same project.
To change the measurement mode for the current project, press the
〈
Mode
〉
button on the
Gauge Setup
menu shown on page 5–2.
The PDA displays the message shown in Figure 5–5.
To maintain the current measurement mode, press
〈
Cancel
〉
.
To change measurement modes, press
〈
OK
〉
. The PDA displays
the setup
Wizard
(see page 3–20). Follow the prompts in the
wizard to select the mode, measurement units, count time, and
so on. Upon completion, the wizard requests a project name
while displaying the currently named project file. To change the
project name, enter a new one, or press
〈
ok
〉
to accept the
current name.
5.
SET
U
P
&
T
ARG
E
T
M
E
NUS
Summary of Contents for Enhanced RoadReader Plus 3451
Page 7: ...Model 3451 vii NOTES...
Page 17: ...xvi NOTES...
Page 29: ...1 12 NOTES...
Page 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 85: ...4 22 NOTES...
Page 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Page 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Page 131: ...6 NOTES...
Page 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Page 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Page 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Page 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Page 153: ...7 22 NOTES...
Page 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Page 174: ...Model 3451 Appendix A 21 NOTES...
Page 175: ......
Page 183: ...Appendix B 8 NOTES...
Page 221: ...Appendix D 6 NOTES...
Page 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 235: ...Appendix G 4 NOTES...