Trox Technik EASYLAB BE-LCD-01 Operating And Installation Manual Download Page 13

BE-LCD-01 Operating and Installation Manual / Leaflet no. M375EV8 (01/2015)

13

5 Operation · Fume cupboard control

Controller faults

Code

Message

Explanation

Actions, remedial measures

H1

24 V 

undervoltage

The supply voltage is too low. The sensors and

actuators are no longer supplied with sufficient

power.

Check the power supply to the

system.

H2

AI 

characteristic

Analogue input signal is outside of the characteristic.

Check the voltage signal or 

sensor. Check the characteristic

parameters stored in the 

controller.

H3

Supportive

flow fan fault

The feedback signal from the supportive flow fan is

missing. The fan has been switched off, or the cable

is no longer connected. The controller automatically

raises the volume flow rate to the value that has been

set for this case.

Check the feedback signal and the

supportive flow fan.

Control panel faults

Code

Message

Explanation

Actions, remedial measures

E1

Connection

lost

Communication between the control panel and the

fume cupboard controller has been interrupted. 

For technical reasons, this text always appears in

English.

Check the cable connection, the

control panel, and the controller.

E2

CP-CRC 

failure

Communication between the control panel and 

the fume cupboard controller is faulty. For technical

reasons, this text always appears in English.

Check the cable connection, the

control panel, and the controller.

E3

CP not 

possible

The control panel cannot be used with the controller

to which it is connected. The control panel must be

connected to a configured fume cupboard controller

or to a controller with active room management

function.

Check the wiring, assignment, and

configuration.

E4

CP memory

A hardware error occurred while accessing the 

internal memory.

If the error occurs again, replace

the control panel.

Operating states, alarm messages, fault displays

Summary of Contents for EASYLAB BE-LCD-01

Page 1: ...ating and Installation Manual Leaflet no M375EV8 01 2015 The art of handling air The art of handling air Operating and Installation Manual EASYLAB Control panel BE LCD 01 for fume cupboard control and...

Page 2: ...Fume cupboard lighting ________________ 10 Overriding operating mode defaults set by the central BMS____________________ 10 Operating states alarm messages fault displays ________________________ 11 5...

Page 3: ...lation manual before starting any work and to comply with it The manual must be given to the facilities manager when handing over the system The facilities manager must include the manual with the sys...

Page 4: ...reign matter if any If the device emits a smell or smoke have it checked by the manufacturer 4 BE LCD 01 Operating and Installation Manual Leaflet no M375EV8 01 2015 2 Safety and correct use Correct u...

Page 5: ...green yellow red with text HIGH and LOW Connection socket for service and commissioning device Acoustic alarm acknowledgement Manual control Selection of operating mode Lighting or other equipment RM...

Page 6: ...tion One button for the fume cupboard lighting One button for manual control Service socket The range of functions of the control panel can be adapted to project specific requirements using the EasyCo...

Page 7: ...udes BE LCD 01 control panel with adapter Connecting cable approx 5 m patch cable blue Operating and installation manual Transport on site If possible take the control panel in the transport packaging...

Page 8: ...of fume cupboard operating states and faults Display texts p 11 onwards The display of additional information can be configured Volume flow rate setpoint and actual values Face velocity actual value o...

Page 9: ...me flow will then automatically be deactivated after the set delay Additional functions Depending on the configuration and operating mode default setting additional displays and function buttons may b...

Page 10: ...ny other operating mode on the control panel This operating mode default setting can be terminated by pressing the manual control button again The fume cupboard controller then resumes the previous op...

Page 11: ...e controller configuration the damper blade will remain open or will be shut shut off SC Automatic sash device Error message relating to the automatic sash device S1 Test function The general test fun...

Page 12: ...f controllers 24 More than 24 controllers have been connected to the communcation cable Reduce the number of controllers C3 Communication cable termination The communication cable has not been termina...

Page 13: ...supportive flow fan Control panel faults Code Message Explanation Actions remedial measures E1 Connection lost Communication between the control panel and the fume cupboard controller has been interru...

Page 14: ...oard 1 Fume cupboard n Supply air Extract air Communication cable The BE LCD 01 control panel is used to display and control the most critical aerodynamic and safety related functions of an EASYLAB co...

Page 15: ...and actual values The display can be adjusted using the configuration software BE LCD 01 Operating and Installation Manual Leaflet no M375EV8 01 2015 15 5 Operation Room control Volume flow rate cont...

Page 16: ...This enables an automatic deactivation of the increased operation after the set delay Note Whether a defaulted operating mode applies to a controller in a room depends on the configuration of the cont...

Page 17: ...an then set any other operating mode using the control panel To resume the previous operating mode from the central BMS press the manual mode button again Manual operation can be set for a limited per...

Page 18: ...ic signal Alarm messages Code State Explanation Actions remedial measures A1 Volume flow rate too high The volume flow rate exceeds the setpoint value Check the actuator and the controller A2 Volume f...

Page 19: ...either no differential pressure controller on the communication cable or the room management function is not active Include a differential pressure controller Activate differential pressure control i...

Page 20: ...ler has been interrupted For technical reasons this text always appears in English Check the cable connection the control panel and the controller E2 CP CRC failure Communication between the control p...

Page 21: ...e for mounting the control panel onto an adapter Danger Danger of electric shock Do not touch any live components Electrical equipment carries a dangerous electrical voltage during operation Only skil...

Page 22: ...nto place See picture on p 23 5 Align the control panel including the cable with the adapter and press it onto the fixing points until it locks into place In the process carefully push the connecting...

Page 23: ...to the right 5 Align the control panel including the cable with the base plate and press it back onto the base plate until it locks into place In the process carefully push the connecting cable back...

Page 24: ...alarm in case the sash is opened beyond the maximum operational sash opening Enabling the function button for Increased Operation Enabling the function button for Reduced operation Enabling the funct...

Reviews: