Lancing Device Cleaning and Disinfecting
• Clean the lancing device immediately after getting any blood on it or if it is dirty.
• Remove lancet from lancing device before cleaning.
• Clean and disinfect the lancing device at least once a week. The lancing device may be
cleaned and disinfected once a day for a 3-year period.
•
Use only PDI Super Sani-Cloth® Wipes
(EPA Reg. No. 9480-4). Other disinfectants have
not been tested and may damage the lancing device.
• PDI Super Sani-Cloth Wipes have been tested on the lancing device for a total of 1,095
cleaning and disinfecting cycles, which is equal to cleaning and disinfecting the lancing
device once a day for a 3-year period.
Important!
Stop using the lancing device and contact Customer Care for assistance, if:
•
markings on lancing device are coming off
•
trigger button is hard to push
•
end cap does not go back on
•
arming barrel does not click when gently pulled back
To Clean the Lancing Device:
1.
Wash hands thoroughly with soap and water.
2.
Thoroughly wipe the entire outside surface of the lancing device with one
Super Sani-Cloth by wiping in circles 3 times. Discard the used cloth.
To Disinfect the Lancing Device:
1.
Always clean the lancing device immediately before disinfecting
(see
instructions above)
.
2.
Remove a fresh Super Sani-Cloth and thoroughly wipe the entire outside surface of
the lancing device by wiping in circles 3 times. Make sure that you go over the entire
lancing device.
3.
Using fresh cloths, if needed,
make sure that all outside surfaces of the lancing
device stay wet for 2 minutes by patting with the cloth
. Discard used cloths.
4.
Let the lancing device air dry thoroughly before using.
5.
Gently pull back the arming barrel and press the trigger button. You will hear a click if
the lancing device is functioning properly.
6.
Wash hands thoroughly when finished.
Call Customer Care at 1-800-803-6025, Mon.–Fri., 8AM–8PM EST