** NOTE ** In the event that the PairMaster and the Remote
Identifier are both connected to an energized cable, only the
PairMaster will give an overvoltage (OL) indication. The OVER-
VOLTAGE INDICATOR LED will not light under these conditions.
In most situations, the Remote Identifier will be connected to
the line first, allowing the user to know immediately if the line is
energized. At the wiring hub (in another office) the user can de-
termine an overvoltage condition by using the PairMaster itself.
FolIowing the initial message at power-up, (i.e. “- -“ ) the
PairMaster tests each pair of the cable sequentially. The pairs
consist of the following conductors: 1/2, 3/6, 4/5, 7/8. The
PASS/FAIL results of each pair are displayed immediately fol-
lowing that pair’s tests.
If the pair has passed all tests, the green LED below the pair
number will light. The digital display will show the number of
the Remote Identifier connected to the other end of the cable.
If the pair fails one or more of the tests, the red LED below the
pair number will light. The digital display will show the FAULT
CODE associated with the failure. In the event of a multiple fail-
ure on a single pair, the highest priority fault will be displayed.
The fault codes are listed in order of priority on page 10.