Specifications
Range Specifications
Accuracies are specified as % r digits at 23
o
C ± 5
o
C < 80% RH
MEGOHMMETER RANGES
Range
Resolution
Accurac
y
Terminal Voltage
200M 250V
0.1M
±3% + 5d 250V + 10% ~ -0%
200M 500V
0.1M
±3% + 5d 500V + 10% ~ -0%
0-1000M 1000VDC
1M
±3% + 5d 1000V + 10% ~ -0%
1000-2000M 1000VDC
1M
±5% + 5d 1000V + 10% ~ -0%
Range
Test Current
Short Circuit
Current
200M 250V
250k (load)
200M 500V
1mA
500k (load)
<1mA
0-1000M 1000VDC
1M
1000-2000M 1000VDC
AC VOLTAGE (40Hz-
400Hz)
Range
Resolution
Accuracy
Input
Impedance
Overload
Protection
750VAC
1V
±1.2% + 10d
10M
750Vrms
www.
.com
1.800.561.8187