Lock Function 

For hands-free operation, use the LOCK feature. 

1. With the test leads connected to the equipment under test, simultaneously press the TEST

and LOCK keys.

2. The LOCK icon”

” will appear on the display. A beeper will sound every 2 seconds to 

indicate that the meter is in Lock mode. 

3. Press the LOCK key to disable the Lock function and end the test.

Notes on IR (Megohmmeter) testing: 

 

5.

The maximum measurement range for the 380260 is 2000MΩ. Frequently, insulation

resistance will exceed this value. When this happens, the display will indicate “1”, meaning

the resistance is very high and the insulation being tested is good.

6.

If the device being tested is highly capacitive, the display will indicate an increasing

resistance value over time. Always wait until the reading has stabilized before recording the

value.

AC/DC Voltage Measurement 

1. Set the Rotary switch to the ACV or DCV position.
2. Connect the red test lead to the V terminal and the black test lead to the COM terminal.
3. Connect the test leads to the circuit under test.
4. Read the voltage value on the LCD.

5.   

Resistance Measurement 

WARNING 

Do not run this test unless ACV = 0. Do not use this mode to check diodes. 

1. Set the Rotary switch to the 200k position.
2. Connect the red test lead to the V terminal and the black lead to the COM terminal.
3. Connect the tips of the test leads to the circuit under test.
4. Read the resistance value on the display.

Low Resistance (Continuity) Measurement 

1.  Set the Rotary switch to the 200Ω

 position. 

 

2. Connect the red test lead to the V terminal and the black lead to the COM terminal.
3. Connect the tips of the test leads to the circuit under test.
4. Read the resistance value on the display. When the resistance of a circuit is less than approx.

40Ω the audible tone will sound.

Auto Power Off 

To conserve battery life, the meter will automatically turn off after approx. 15 minutes of non-use. To 

turn the meter back on, turn the rotary switch to OFF, then to the desired function. 

Hold Function 

The hold function freezes the reading in the display. Press the HOLD key momentarily to activate or 

to exit the hold function 

Backlight 

Press the key for to turn on the display backlight function. The backlight will automatically turn off 

after 15 seconds. 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for MG430

Page 1: ...measurements Circuit connections must not be touched during a test Use extreme caution when working near bare conductors and bus bars Accidental contact with conductors could result in electrical shoc...

Page 2: ...eplaced before using the meter Failure to do so could result in damage to equipment and electrical shock Insulation Resistance Measurements Megohmmeter Tests Warning Do not perform Insulation Resistan...

Page 3: ...he voltage value on the LCD 5 Resistance Measurement WARNING Do not run this test unless ACV 0 Do not use this mode to check diodes 1 Set the Rotary switch to the 200k position 2 Connect the red test...

Page 4: ...is indicated the motor and other components should be checked individually If the motor is disconnected at the motor terminals connect one meter lead to the grounded motor housing and the other lead t...

Page 5: ...0 200M 500V 0 1M 3 5d 500V 10 0 0 1000M 1000VDC 1M 3 5d 1000V 10 0 1000 2000M 1000VDC 1M 5 5d 1000V 10 0 Range Test Current Short Circuit Current 200M 250V 250k load 200M 500V 1mA 500k load 1mA 0 100...

Page 6: ...ut Impedance Overload Protection 1000VDC 1V 0 8 3d 10M 1000Vrms RESISTANCE Range Resolution Accuracy Max Open Circuit Overload Voltage Protection 200 0 1 1 2d 4 5V 250Vrms 200k 0 1k 1 2d 3V 250Vrms ww...

Page 7: ...ymbol displayed when battery voltage is low Environmental Installation category III Pollution Degree 2 Altitude up to 2000 meters Indoor use only Power source Six 6 1 5 AA batteries Fuse 250mA 600V ce...

Page 8: ...l purchaser of these goods for use Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of 1 one year fro...

Reviews: