33% Calibration
1.
Plug the sensor probe into the 33% salt bottle.
2.
Long press the
/ESC, MEM/
and
/SET buttons
simultaneously to enter calibration mode.
3.
In calibration mode, “CAL” and the current value (32.8% if at
25°C) will blink on the LCD.
4.
The meter is now calibrating, and will finish in about 60
minutes, when the current value stops blinking. “CAL” will
continue to blink on the display.
5.
To abort the calibration, power the meter OFF at any time.
75% calibration
1.
After the 33% calibration, plug the sensor probe into a 75% salt
bottle.
2.
Long press the MXMN/AV/
button to enter 75% calibration.
3.
“CAL” and current value (75.3% if at 25°C) will blink on the LCD.
4.
The meter is now calibrating. The calibration is complete in
about 60 minutes. The values will stop flashing and the meter
powers OFF automatically when calibration is complete.
5.
To abort the calibration, power OFF the meter at any time.