background image

20

21

CONTACT INFORMATION

EMC Guidance

Contact Information

For more information about our products, please visit www.transtek.cn.you can get customer service, 

usual problems and customer download, transtek will serve you anytime.

Authorized European Representative:

Company:

 MDSS - Medical Device Safety Service GmbH

Address:

 Schiffgraben 41, 30175 Hannover, Germany

Manufactured by:

GUANGDONG TRANSTEK MEDICAL ELECTRONICS CO., LTD

Company:

GUANGDONG TRANSTEK MEDICAL ELECTRONICS CO., LTD

Address:

Address: Zone A, No.105 ,Dongli Road, Torch Development District, 

Zhongshan,528437,Guangdong,China  

Complied European Standards List

EMC GUIDANCE

Risk management

Labeling

User manual

General 

Requirements for 

Safety

Electromagnetic 

compatibility

Performance 

requirements

Clinical investigation

Usability

Software life-cycle 

processes

EN ISO 14971:2012 Medical devices - Application of risk 

management to medical devices

EN 980:2008 Symbols for use in the labelling of medical 

devices

EN 1041:2008 Information supplied by the manufacturer of 

medical devices

EN 60601-1:2006 Medical electrical equipment - Part 1: 

General requirements for basic safety and essential 

performance

EN 60601-1-11:2010 Medical electrical equipment - Part 

1-11: General requirements for basic safety and essential 

performance - Collateral standard: Requirements for medical 

electrical equipment and medical electrical systems used in 

the home healthcare environment

EN 60601-1-2:2007 Medical electrical equipment - Part 

1-2: General requirements for basic safety and essential 

performance - Collateral standard: Electromagnetic 

compatibility - Requirements and tests

EN ISO 81060-1:2012 Non-invasive sphygmomanometers 

- Part 1: Requirements and test methods for non-automated 

measurement type

EN 1060-3:1997+A2:2009 Non-invasive 

sphygmomanometers - Part 3: Supplementary requirements 

for electro-mechanical blood pressure measuring systems

EN 1060-4:2004 Non-invasive sphygmomanometers - Part 

4: Test procedures to determine the overall system accuracy 

of automated non-invasive sphygmomanometers

EN 60601-1-6:2010 Medical electrical equipment - Part 

1-6: General requirements for basic safety and essential 

performance - Collateral standard: Usability

EN 62366:2008 Medical devices - Application of usability 

engineering to medical devices

EN 62304:2006/AC: 2008  Medical device software - 

Software life cycle processes

Table 1 Guidance and MANUFACTURER’s declaration – ELECTROMAGNETIC 

EMISSIONS- for all ME EQUIPMENT and ME SYSTEMS 

Guidance and manufacturer’s declaration – electromagnetic emissions

RF emissions 
CISPR 11

Group 1 

Class B 

Not applicable 

Not applicable 

Compliance

The device is intended for use in the electromagnetic environment specified

 below. The customer or the user of the device should assure that it is used in 

such an environment.

Harmonic emissions 
IEC 61000-3-2

Voltage fluctuations/ 
flicker emissions IEC 
61000-3-3  

RF emissions 
CISPR 11

Emissions test

Electromagnetic environment - guidance

The device uses RF energy only for its 

internal function. Therefore, its RF

emissions are very low and are not likely

to cause any interference in nearby 

electronic equipment.

1) * This product needs special precautions regarding EMC and needs to be 

installed and put into service according to the EMC information provided, and 

this unit can be affected by portable and mobile RF communications equipment.

2) * Do not use a mobile phone or other devices that emit electromagnetic 

fields, near the unit. This may result in incorrect operation of the unit.

3) * Caution: This unit has been thoroughly tested and inspected to assure 

proper performance and operation!

4) * Caution: This machine should not be used adjacent to or stacked with other 

equipment and that if adjacent or stacked use is necessary, this machine should 

be observed to verify normal operation in the configuration in which it will be 

used.

The device is suitable for use in all establishments, 

including domestic establishments and those directly

connected to the public low-voltage power supply

network that supplies buildings used for domestic

purposes.

Summary of Contents for TMB-986

Page 1: ...ual carefully and thoroughly so as to ensure the safe usage of this product and keep the manual well for further refer ence in case you have problems GUANGDONG TRANSTEK MEDICAL ELECTRONICS CO LTD Address Zone A No 105 Dongli Road Torch Development District Zhongshan 528437 Guangdong China EC REP IFU TRANSTEK TMB986EN 2016_02 TMB 986 英文说明书 A0 印色 单黑 材质 80G书纸 尺寸 170 120mm ...

Page 2: ...ATION FOR USER 14 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 16 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why does my blood pressure fluctuate throughout the day Why the blood pressure I get from the hospital is different from home If the result is the same if measuring on the right arm TROUBLESHOOTING 18 SPECIFICATIONS 19 CONT...

Page 3: ...ent errors please avoid the condition of strong electromagnetic field radiated interference signal or electrical fast transient burst signal The user must check that the equipment functions safely and see that it is in proper working condition before being used This device is contraindicated for any female who may be suspected of or is pregnant Besides provided inaccurate readings the affects of t...

Page 4: ...ime hour minute Currently time Memory If M shows the displayed measurement values is from the memory mmHg Measurement Unit of the blood pressure 1mmHg Low battery Batteries are low and need to be replaced Shocking remainding Shocking will result in inaccurate Average The average of blood pressure Recalling The erecords will be showed Irregular heartbeat Irregular heartbeat User A The User A is cho...

Page 5: ...e off the battery cover 2 Install the batteries by matching the correct polarity as shown 3 Replace the cover Remove batteries if the device is not likely to be used for some time The old batteries are harmful to the environment so please disposal with other daily trash Remove the old batteries from the device and follow your local recycling guidelines CAUTION Replace the batteries whenever the be...

Page 6: ...Repeat step 2 and 3 to set the HOUR and MINUTE START STOP START STOP START STOP START STOP START STOP START STOP START STOP START STOP BEFORE YOU START BEFORE YOU START 6 After confirming the minute the LCD will display dOnE and the monitor will shut off ...

Page 7: ...ke daily measurements at approximately the same time on the same arm or as directed by a physician Tie the Cuff 1 Tie the cuff on your upper arm the position the tube off center toward the inner side of arm in line with the little finger 2 The cuff should be sung but not too tight You should be able to insert one fingers between the cuff and your arm 3 Sit comfortably with your arm resting on a fl...

Page 8: ...s 1 Please choose the user A or user B when the blood pressure monitor turns off and press the MEM to show the average of the last 3 records The sign of AVG will show in the right corner 2 Press the MEM or SET to get the record you want The most recent record 1 is shown first Each new measurement is assigned to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so o...

Page 9: ... or moving your fingers When you want to discharge urine 15 14 Maintenance In order to get the best performance please follow the below instructions Put in a dry place and avoid the sunshine Avoid the intense shaking and collision Using the wet clothing to remove the dirt Avoid touching water clean it with a dry cloth in case Avoid the dusty and unstable temperature environment Avoid washing the c...

Page 10: ...y a physician could tell whether your blood pressure value has reached a dangerous point The blood pressure classification published by World Health Organization WHO and International Society of Hypertension ISH in 1999 is as follows SYS DIA 120 80 120 129 80 84 130 139 85 89 140 159 90 99 160 179 100 109 180 110 Level Blood Pressure mm Hg Optimal Normal High normal Mild Moderate Severe CAUTION 1 ...

Page 11: ...shows on the display A calibration error occurred Retake the measurement If the problem persists contact the retailer or our customer service department for further assistance Refer to the warranty for contact information and return instructions E21 shows The treatment of the measurement failed Relax for a moment and then measure again Display is dim or shows Power supply Battery powered mode 4 AA...

Page 12: ...s Part 4 Test procedures to determine the overall system accuracy of automated non invasive sphygmomanometers EN 60601 1 6 2010 Medical electrical equipment Part 1 6 General requirements for basic safety and essential performance Collateral standard Usability EN 62366 2008 Medical devices Application of usability engineering to medical devices EN 62304 2006 AC 2008 Medical device software Software...

Page 13: ...NG Guidance and manufacturer s declaration electromagnetic immunity IMMUNITY test Not applicable 3 V m Compliance level IEC 60601 TEST LEVEL Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 Vrms 150 kHz to 80 MHz Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the device including cables than...

Page 14: ...ters and the device as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of th...

Reviews: