WARNING: No modification of this equipment is allowed.
18
19
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the products not operating as you think it
should, check here before arranging for servicing.
PROBLEM SYMPTOM CHECK THIS
REMEDY
No power
Low
batteries
Error
message
Display will not
light up.
Batteries are exhausted. Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Batteries are low.
E 1 shows
The cuff is not secure.
Refasten the cuff and then
measure again.
E 2 shows
The cuff is very tight
Refasten the cuff and then
measure again.
E 3 shows
The pressure of the
cuff is excess.
Relax for a moment
and then measure again.
E10 or E11
shows
The monitor detected
motion while measuring.
Movement can affect the
measurement.Relax for a
moment and then
measure again.
E20 shows
The measurement
process does not detect
the pulse signal.
Loosen the clothing on the
arm and then measure
again
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
E21 shows
The treatment of the
measurement failed.
Relax for a moment and
then measure again.
Display is dim or
shows
Power supply
Battery powered mode:
4*AAA batteries,6V
Display mode
Digital LCD display,
V.A.128mm*50mm
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the upper arm
About 22cm~42cm
Net Weight
Approx.320g(Excluding the dry cells)
External dimensions
Attachment
Approx.182mm*100mm*39mm
4*AAA batteries,user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
Protection against
ingress of water
IP21
Software Version
V01
Temperature:5
℃
to 40
℃
Atmospheric pressure: 86kPa to 106kPa
Temperature:-20
℃
-60
℃
Relative Humidity: 10%RH-93%RH
Atmospheric Pressure: 50kPa-106 kPa
TROUBLESHOOTING
SPECIFICATIONS
Rated cuff pressure:
0mmHg~300mmHg
Measurement pressure:
40mmHg-230mmHg
pulse value: (40-199) beat/minute
Pressure:
5
℃
-40
℃
within±3mmHg
pulse value:±5%
Relative humidity ≤85%RH