background image

CAUTION

* When using this device, please pay attention to the following situation which may interrupt blood flow and 

influence blood circulation of the patient, thus cause harmful injury to the patient: connection tubing kinking 

too frequent and consecutive multiple measurements; the application of the cuff and its pressurization on 

any arm where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present; inflating the 

cuff on the side of a mastectomy.

* Warning: Do not apply the cuff over a wound;otherwise it can cause further injury.

*Do not inflate the cuff on the same limb which other monitoring ME equipment is applied around 

simultaneously, because this could cause temporary loss of function of those simultaneously-used 

monitoring ME equipment.

*On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff 

immediately. Prolonged high pressure (cuff pressure 

 300mmHg or constant pressure 

 15mmHg for 

more than 3 minutes) applied to the arm may lead to an ecchymosis.

*Please check that operation of the device does not result in prolonged impairment of patient blood 

circulation.

* When measuring, please avoid compression or restriction of the connection tubing.

* The device cannot be used with HF surgical equipment at the same time.

* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER is clinically 

investigated according to the requirements of ISO 81060-2:2013.

* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the manufacturer.

* This device is contraindicated for any female who may be suspected of, or is pregnant. Besides providing 

inaccurate readings, the effects of this device on the fetus are unknown.

* Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.

* This unit is not suitable for continuous monitoring during medical emergencies or operations.Otherwise, 

the patient’s arm and fingers will become anaesthetic, swollen and even purple due to a lack of blood.

* When not in use, store the device with the adapter in a dry room and protect it against extreme moisture, 

heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.

* This device may be used only for the purpose described in this booklet. The manufacturer cannot be held 

liable for damage caused by incorrect application.

*This device comprises sensitive components and must be treated with caution. Observe the storage and 

operating conditions described in this booklet.

* The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable 

anesthetic mixture with air of with oxygen or nitrous oxide.

* Warning: No servicing/maintenance while the ME equipment is in use.

* The patient is an intended operator.

* The patient can measure, transmit data and change batteries under normal circumstances and maintain 

the device and its accessories according to the user manual.

* To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated 

interference signal or electrical fast transient/burst signal.

*The blood pressure monitor, its adaptor, and the cuff are suitable for use within the patient environment. If 

you are allergic to polyester, nylon or plastic, please don't use this device. 

* During use, the patient will be in contact with the cuff. The materials of the cuff have been tested and 

found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any 

potential sensization or irritation reaction.

* Adaptor is specified as a part of ME EQUIPMENT.

* If you experience discomfort during a measurement, such as pain in the arm or other complaints, press 

the START/STOP button to release the air immediately from the cuff. Loosen the cuff and remove it from 

your arm.

* Do not wash the cuff in a washing machine or dishwasher!

* The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The 

typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after maintenance and 

repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air 

leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the local 

guidelines.

CAUTION

 * Manufacturer will make available on request circuit diagrams, component part lists, descriptions, 

calibration instructions,etc., to assist to service personnel in parts repair.

* The plug/adapter plug pins insulates the device from the main supply. Do not position the device in a 

position where it is difficult to disconnect from the supply mains to safely terminate operation of ME 

equipment.

* The operator shall not touch output of batteries /adapter and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the soft cloth to clean 

the whole unit before and after use. Don’t use any abrasive or volatile cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, please contact the 

SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of 

malfunctions. The device must only be serviced, repaired and opened by individuals at authorized 

sales/service centers.

* Please report to Transtek if any unexpected operation or events occur.

* Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of small 

parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive length.

* At least 30 min required for ME equipment to warm from the minimum storage temperature between 

uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the 

maximum storage temperature between uses until it is ready for intended use.

* This equipment needs to be installed and put into service in accordance with the information provided in 

the ACCOMPANYING DOCUMENTS;

* Wireless communications equipment such as wireless home network devices, mobile phones, cordless 

telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a 

distance d away from the equipment. The distance d is calculated by the MANUFACTURER from the 80 

MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.

* Please use ACCESSORIES and detachable parts specified/ authorised by MANUFACTURER. 

Otherwise, it may cause damage to the unit or danger to the user/patients.

* There is no luer lock connectors are used in the construction of tubing. If there has,  there is a possibility 

that they might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a 

blood vessel.

* Please use the device under the environment which is provided in the user manual. Otherwise, the 

performance and lifetime of the device will be impacted and reduced.

4

5

INTRODUCTION

INTRODUCTION

Summary of Contents for TMB-1776-B

Page 1: ...B 1776 B Please read the user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems Guangdong Transte...

Page 2: ...Delete the Records INFORMATION FOR USER 17 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood pressure classificati...

Page 3: ...n for treatment advice If you are taking medication consult your physician to determine the most appropriate time to measure your blood pressure Never change a prescribed medication without consulting...

Page 4: ...and ISO 10993 10 2010 It will not cause any potential sensization or irritation reaction Adaptor is specified as a part of ME EQUIPMENT If you experience discomfort during a measurement such as pain...

Page 5: ...g measurement User A User B Start measurement and save the measuring results for User A User B BLJ06L060100P U 7 6 INTRODUCTION INTRODUCTION Monitor Components List 1 Blood Pressure Monitor TMB 1776 B...

Page 6: ...s not likely to be used for some time Worn batteries are harmful to the environment Do not dispose with daily garbage Remove the old batteries from the device following your local recycling guidelines...

Page 7: ...onE and then turn off Select the User ID Before you start the measurement please select the desired user ID first When the blood pressure monitor is off press USER button the user ID will blink Then p...

Page 8: ...e 5 6 deep breaths Rest for 5 minutes before first measuring Wait at least 3 minutes between measurements This allows your blood circulation to recover The patient must relax as much as possible and d...

Page 9: ...our mobile device the measurement data will be automatically transmitted to your mobile device via Bluetooth 2 The symbol will disappear after successful data transmission and you may check your perso...

Page 10: ...onding date is January 1st The date and time of the record will be shown alternately If you did not get the correct measurement you can delete the result for the selected user by following steps below...

Page 11: ...button to confirm deleting when it shows dELyES then the monitor with turn off when it shows User donE 4 If there is no record press MEM button the right display will be shown Tips Press START STOP b...

Page 12: ...ressure classification The chart on the right is the standard blood pressure classifi cation published by American Heart Association AHA Please consult a physician if your measuring result falls outsi...

Page 13: ...ssure fluctuate throughout the day Is the result the same if measuring on the right arm Why do I get a different blood pressure at home compared to the hospital 1 Individual blood pressure varies mult...

Page 14: ...value 5 Digital LCD display V A 80mm 58 5mm Approx 140 4mm 110 4mm 64 8mm 4 AA batteries user manual AC adaptor carry bag IP21 It means the device could protected against solid foreign objects of 12...

Page 15: ...Part 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labelling and information to be supplied Part 1...

Page 16: ...sions RF emissions CISPR 11 Group 1 Class B Class A Comply Compliance Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 RF emissions CISPR 11 Emissions test Guidanc...

Page 17: ...430 470 GMRS 460 FRS 460 FM c 5kHz deviation 1kHz sine 2 0 3 28 710 704 787 745 780 LTE Band 13 17 Pulse modulation b 217Hz 0 2 0 3 9 810 870 930 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE B...

Reviews: