background image

10

11

Problem

Root Cause

Solution

Incorrect posture

Abnormal 

measuring results:

- Too high; OR

- Too low; OR

- Huge difference 

between two recent 

measurement.

Please step on the platform 

barefooted and stand still.

The device is located on 

the soft ground such as a 

carpet OR on a rugged 

surface.

Please place the device on a 

flat, hard surface.

Your feet are too dry.

Abnormal 

measuring results:

- Too high; OR

- Too low; OR

- Huge difference 

between two recent 

measurement.

No display on 

LCD when the 

device powers on.

The device 

powers off.

Wipe your feet with a damp 

cloth, keeping them slightly damp 

when starting measurement.

Batteries not yet installed.

Worn batteries.

Low battery.

Install the batteries.

(Please refer to Insert the 

Batteries)

Replace all four batteries at the 

same time. Please purchase the 

authorized batteries for 

replacement.

Replace all four batteries at the 

same time. Please purchase the 

authorized batteries for 

replacement.

When Measuring ...

Specifications

About the Accuracy of This Product

This product passes strict inspection before delivery and therefore its accuracy is guaranteed by the manufacturer. 

Please refer to the above table for the descriptions on accuracy.

This product is specially designed for body fat analysis as well as weight measurement. It should NOT be used by 

anyone during the process of transaction for verification of goods’ weight.

Body Fat Analyzer (GBF-1717-W)

0.1 Ib

6V (Four AAA-size Batteries)

Product Name

Dimension
Net Weight

Display

Digital LCD 

Measurement Unit

Pound

Measurement Range

397Ib

Division

Accuracy

50kg: ±0.2kg;

150kg: ±0.3kg;

Working Environment

Storage Environment

Power Source

Auto-ON

SENSE ON technology

Auto-OFF

The scale will turn off after about 10s if there is no operation.

Accessories

1. Four AAA Size Batteries

2. User Manual

100kg: ±0.2kg;

180kg: ±0.4kg;

Mode of Operation

Continuous Operation

Temperature: 5

 to 40

Temperature: -20

 to 60

Relative Humidity: 

≤90% 

RH

Atmospheric pressure: 86kPa to 106kPa

Atmospheric pressure: 50kPa to 106kPa

Relative Humidity: 

≤90% 

RH

Scale: 321x321x25.5mm (Approximately)

Approximately 2.08kg ( Including the dry cells)

Wifi Frequency Range

5.180GHZ~5.825GHZ

Wifi Output Power

14~17dBm

Wifi Module No

Wifi Supply Voltage

WG222

3.3V±10%

Software Version

V1.06

Degree of protection

Type BF applied part

Troubleshooting

Specifications

Protection against 

ingress of water

IP21

Summary of Contents for GBF-1717-W

Page 1: ...ensure the safe usage of this product and keep the manual well for further reference in case you have problems Thank you very much for selecting the Transtek Glass Body Fat Analyzer GBF 1717 W TEL 86 760 88282982 www transtek cn ZHONGSHAN TRANSTEK ELECTRONICS CO LTD No 23 Jin an Road Minzhong Zhongshan 528441 Guangdong China ...

Page 2: ...y Measurement Troubleshooting Error Prompt When Data Transmitting When Measuring Specifications Maintenance Maintenance Warranty FCC Regulations Appendix Health Tips About Body Fat Health Tips About Body Water Health Tips About Muscle Mass Health Tips About Bone Mass Warning EMC Guidance 2 3 4 4 5 6 7 9 9 10 11 12 12 13 14 14 14 15 16 16 Data Management Data Transmission 8 Users Management 7 ...

Page 3: ...ping Please keep the device out of reach of infants children or pets since inhalation or swallowing of small parts is dangerous or even fatal Please don t use the device out of lifetime any more because degraded sensors and electrodes or loosened electrodes can degrade performance or cause other problems If you have an allergy to stainless steel please avoid to contact to the electrodes of the dev...

Page 4: ...7 W 2 Four AAA size Batteries 1 5V per each 3 User Manual Device Components Battery Compartment Sensor Data pending to transmit to smartphone Pound Successful APP Connection Touch Key LCD Display Successful analysis User Key ID1 User Key ID2 User Key ID3 User Key ID4 ...

Page 5: ...nces and recycling instructions regarding disposal of the worn batteries With original SENSE ON patent technology the scale will automatically switch on as you step on the platform barefooted When your scale is successfully connected to your smartphone and the WiFi connection is ON it will process data transmission automatically Please refer to Data Transmission for more details Daily Measurement ...

Page 6: ...ssful APP Connection Data Management Error Prompt Error Description Solution Overload The device will power off Stop using this scale for measurement Replace all four batteries at the same time Please purchase the authorized batteries for replacement Stand on the scale and measure again Measurement error Make sure the Wifi connection is successful Data transmission failed Low Battery When the symb...

Page 7: ...uracy is guaranteed by the manufacturer Please refer to the above table for the descriptions on accuracy This product is specially designed for body fat analysis as well as weight measurement It should NOT be used by anyone during the process of transaction for verification of goods weight Body Fat Analyzer GBF 1717 W 0 1 Ib 6V Four AAA size Batteries Product Name Dimension Net Weight Display Digi...

Page 8: ...accordance with the instructions supplied This device complies with FCC radiation exposure limits set forth for an uncontrolled environment FCC Regulations When carrying out usual maintenance please ensure practice of the following Do s and Don ts Maintenance DO use a dry soft cloth to wipe the dust DO use a wet soft cloth dipped into water and wrung out to wipe the dirt Then use a dry soft cloth ...

Page 9: ...larly through perspiration The combination of your weight and fat measurement could appear to be normal but your body hydration level could be insufficient for healthy living Health Tips About Muscle Mass According to the American College of Sports Medicine ACSM lean muscle mass may decrease by nearly 50 percent between the age of 20 and 90 If you do nothing with it you re losing muscle and increa...

Page 10: ...of the performance of this equipment could result Guidance and manufacturer s declaration electromagnetic emissions and Immunity Table 1 Guidance and manufacturer s declaration electromagnetic emissions Emissions test Compliance RF emissions CISPR 11 Group 1 RF emissions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applic...

Page 11: ...390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 2 0 3 28 870 930 1720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 217 Hz 2 0 3 28 1845 1...

Reviews: