* Please check that the operation of the device do not result in prolonged impairment of patient blood
circulation.
*
Warning:
On the rare occasion of a fault causing the cuff to remain fully inflated during
measurement, loosen and remove the cuff immediately. Prolonged high pressure applied to the arm
(cuff pressure >300 mmHg or constant pressure >15 mmHg for more than 3 minutes) might lead to
bruising and discolored skin.
*
Warning:
Do not use this device with high-frequency (HF) surgical equipment at the same time.
*
Warning:
This device is not AP/APG equipment. Do not use the device where flammable
anesthetic are present, or in environments mixture with air of with oxygen or nitrous oxide.
* The device contains sensitive electronic components.To avoid measurement errors, avoid taking blood
pressure measurements near a strong electromagnetic field radiated interference signal or electrical fast
transient/burst signal.
* Wireless communication equipment, such as wireless home network devices, mobile phones, cordless
telephones and their base stations, walkie-talkies may cause interference that may affect the accuracy of
measurements. A minimum distance of 1 foot (30 cm) should be kept from such devices during a
measurement.
* You can use this device to take your own measurement, no third-party operator is required.
* Please use the device under the environment which is provided in the user manual. Otherwise, the
performance and lifetime of the device will be impacted and reduced.
* The device may require up to 30 minutes to warm up / cool down from the minimum / maximum
storage temperature before it is ready for use.
*
Warning:
Excessive cuff tube lengths could cause strangulation if you don't manage them
properly.
*
Warning:
Do not touch output of the batteries/adapter and the user simultaneously.
* Adapter is specified as a part of ME EQUIPMENT.
* Warning: The power cord is considered the disconnect device for isolating this equipment from supply
mains. Do not position the equipment so that it is difficult to reach or disconnect.
* The blood pressure monitor, its adapter, and the cuff are suitable for use within the patient
environment.
*
Warning:
Do not use this device if you are allergic to polyester, nylon, or plastic.
*
Warning:
Only use accessories approved by manufacturer. Using unapproved accessories might cause
damage to the unit and injure users.
*
Warning:
If you experience discomfort during a measurement, such as pain in the arm or other
complaints, press the Power button immediately to release the air from the cuff.
* No calibration is required within two years of reliable service.
* Do not attempt to repair the unit yourself if it malfunctions. Only have repairs carried out by authorized
service centers.
* At the request of authorized service personnel, circuit diagrams, component part lists,
descriptions, and calibration procedures will be made available by the manufacturer or
distributor.
* This device is intended for indoor use.
* This device is portable, but it is not intended for use during patient transport.
* This device is not suitable for continuous monitoring during medical emergencies or operations.
* This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not
intended for use on extremities other than the arm, or for any purpose other than obtaining a blood
pressure measurement.
* This device is for adults. Do not use this device on neonates or infants. Do not use it on children unless
otherwise instructed by a medical professional.
* The effectiveness of this device has not been established for use:
- on users in pregnant, including pre-eclamptic, patients,
- on users with implanted, electrical devices, such as a pacemaker or defibrillator,
- on users with common arrhythmias such as atria ventricular, premature beats or atrial fibrillation,
- on users with peripheral arterial disease,
- on users undergoing intravascular therapy, or with arteriovenous (AV) shunt.
Consult a medical professional before use.
* Do not use this device for diagnosis or treatment of any health problem or disease. Contact your
physician if you have or suspect any medical problem. Do not change your medications without the
advice of your physician or health care professional.
* If you are taking medication, consult your physician to determine the proper time to measure your
blood pressure.
* This device may be used only for the intended use described in this manual, the manufacturer shall
have no liability for any incidental, consequential, or special damages caused by misuse or abuse.
* Report any unexpected operation or events to the manufacturer.
* Do not apply the cuff on an arm that has an intravenous drip or a blood transfusion attached.
*
Warning:
Do not kink, fold, stretch, compress, or otherwise deform the tube during measuring, as the
cuff pressure might continuously increase, which could prevent blood flow and result injury.
* Warning: Taking blood pressure measurements too frequently could disrupt blood circulation and
cause injuries.
*
Warning:
Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site
frequently for irritation.
*
Warning:
Do not place the cuff on the arm of a person whose arteries or veins are undergoing medical
treatment, i.e. intra-vascular access or intra-vascular therapy or an arteriovenous (A-V) shunt, which
could disrupt blood circulation and cause injuries.
* Do not place the cuff on the arm on the same side of a mastectomy (especially when lymph nodes
have been removed). it is recommended to take measurements on the unaffected side.
* Do not wrap the cuff on the same arm to which another monitoring device is applied. One or both
devices could temporarily stop functioning if you try to use them at the same time.