TRANSTEK Carbon Health CH-B1-11-AA User Manual Download Page 4

* Please check that the operation of the device do not result in prolonged impairment of patient blood 
circulation.

Warning: 

On the rare occasion of a fault causing the cuff to remain fully inflated during

   measurement, loosen and remove the cuff immediately. Prolonged high pressure applied to the arm 
(cuff pressure >300 mmHg or constant pressure >15 mmHg for more than 3 minutes) might lead to 
bruising and discolored skin.

Warning: 

Do not use this device with high-frequency (HF) surgical equipment at the same time.

Warning: 

This device is not AP/APG equipment. Do not use the device where flammable

   anesthetic are present, or in environments mixture with air of with oxygen or nitrous oxide.
* The device contains sensitive electronic components.To avoid measurement errors, avoid taking blood 
pressure measurements near a strong electromagnetic field radiated interference signal or electrical fast 
transient/burst signal.
* Wireless communication equipment, such as wireless home network devices, mobile phones, cordless 
telephones and their base stations, walkie-talkies may cause interference that may affect the accuracy of 
measurements. A minimum distance of 1 foot (30 cm) should be kept from such devices during a 
measurement.
* You can use this device to take your own measurement, no third-party operator is required.
* Please use the device under the environment which is provided in the user manual. Otherwise, the 
performance and lifetime of the device will be impacted and reduced.
* The device may require up to 30 minutes to warm up / cool down from the minimum / maximum 
storage temperature before it is ready for use.

Warning: 

Excessive cuff tube lengths could cause strangulation if you don't manage them

   properly.

Warning: 

Do not touch output of the batteries/adapter and the user simultaneously.

* Adapter is specified as a part of ME EQUIPMENT.
* Warning: The power cord is considered the disconnect device for isolating this equipment from supply 
mains. Do not position the equipment so that it is difficult to reach or disconnect.
* The blood pressure monitor, its adapter, and the cuff are suitable for use within the patient 
environment.

Warning: 

Do not use this device if you are allergic to polyester, nylon, or plastic.

Warning: 

Only use accessories approved by manufacturer. Using unapproved accessories might cause 

damage to the unit and injure users.

Warning: 

If you experience discomfort during a measurement, such as pain in the arm or other 

complaints, press the Power button immediately to release the air from the cuff.
* No calibration is required within two years of reliable service.
* Do not attempt to repair the unit yourself if it malfunctions. Only have repairs carried out by authorized 
service centers.
* At the request of authorized service personnel, circuit diagrams, component part lists,
   descriptions, and calibration procedures will be made available by the manufacturer or
   distributor.

* This device is intended for indoor use.
* This device is portable, but it is not intended for use during patient transport.
* This device is not suitable for continuous monitoring during medical emergencies or operations.
* This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not 
intended for use on extremities other than the arm, or for any purpose other than obtaining a blood 
pressure measurement.
* This device is for adults. Do not use this device on neonates or infants. Do not use it on children unless 
otherwise instructed by a medical professional.
* The effectiveness of this device has not been established for use:
   - on users in pregnant, including pre-eclamptic, patients,
   - on users with implanted, electrical devices, such as a pacemaker or defibrillator,
   - on users with common arrhythmias such as atria ventricular, premature beats or atrial fibrillation,
   - on users with peripheral arterial disease,
   - on users undergoing intravascular therapy, or with arteriovenous (AV) shunt.
   Consult a medical professional before use.
* Do not use this device for diagnosis or treatment of any health problem or disease. Contact your 
physician if you have or suspect any medical problem. Do not change your medications without the 
advice of your physician or health care professional.
* If you are taking medication, consult your physician to determine the proper time to measure your 
blood pressure.
* This device may be used only for the intended use described in this manual, the manufacturer shall 
have no liability for any incidental, consequential, or special damages caused by misuse or abuse.
* Report any unexpected operation or events to the manufacturer.
* Do not apply the cuff on an arm that has an intravenous drip or a blood transfusion attached.

Warning:

 Do not kink, fold, stretch, compress, or otherwise deform the tube during measuring, as the 

cuff pressure might continuously increase, which could prevent blood flow and result injury.
* Warning: Taking blood pressure measurements too frequently could disrupt blood circulation and 
cause injuries.

Warning:

 Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site 

frequently for irritation.

Warning:

 Do not place the cuff on the arm of a person whose arteries or veins are undergoing medical 

treatment, i.e. intra-vascular access or intra-vascular therapy or an arteriovenous (A-V) shunt, which 
could disrupt blood circulation and cause injuries.
* Do not place the cuff on the arm on the same side of a mastectomy (especially when lymph nodes 
have been removed). it is recommended to take measurements on the unaffected side.
* Do not wrap the cuff on the same arm to which another monitoring device is applied. One or both 
devices could temporarily stop functioning if you try to use them at the same time.

CAUTION

CAUTION

INTRODUCTION

INTRODUCTION

6

5

Summary of Contents for Carbon Health CH-B1-11-AA

Page 1: ...1 0 Guangdong Transtek Medical Electronics Co Ltd Zone A No 105 Dongli Road Torch Development District 528437 Zhongshan Guangdong China User Manual Blood Pressure Monitor Bluetooth Connected Model CH...

Page 2: ...pressure and diastolic pressure What is the standard blood pressure classi cation Why does my blood pressure fluctuate throughout the day Why do i get a different blood pressure at home compared to t...

Page 3: ...e and also pulse rate This Pro Cellular Blood Pressure Monitor CH B2 11 AA is a digital monitor intended for use in measuring blood pressure and heartbeat rate with arm circumference ranging from 22 t...

Page 4: ...s component part lists descriptions and calibration procedures will be made available by the manufacturer or distributor This device is intended for indoor use This device is portable but it is not in...

Page 5: ...ation and air leakage testing at least at 50 mmHg and 200 mmHg Warning Do not use the device while under maintenance or being serviced Store your device cuff and adapter in a clean and dry place prote...

Page 6: ...ou nish the measurement 5V 1A Replace the batteries whenever the below happens The low battery symbol appears on the display When any button is pressed and nothing is displayed on the screen BATTERY C...

Page 7: ...oth and the Carbon Health App Make sure both are ON when pair up is proceeding 3 When the monitor is OFF press and hold the START STOP button unitl the symbol appears on the LCD and blink indicating p...

Page 8: ...o insert one nger between the cuff and your arm 2 3 cm 1 inch MEASUREMENT Interference may occur in the vicinity of equipment marked with the following symbol And CH B1 11 AA may interfer the vicinity...

Page 9: ...nd the oldest record will drops from the list after the measurement Caution You can either record your measurement results manually along with the time of measurement or transfer them electronically t...

Page 10: ...d avoid the sunshine Avoid intense shaking and collisions Use a damp cloth to remove dirt Avoid washing the cuff Avoid dusty and unstable temperature environments Within 20 minutes after taking a bath...

Page 11: ...e Please contact a physician if your measurement falls out of the range Please note that only a physician can tell whether your blood pressure value has reached a dangerous point The following chart i...

Page 12: ...for 4 5 minutes to calm down If the cuff is tied properly If the cuff is too tight or too loose If the cuff is tied on the upper arm If you feel anxious 22 21 PROBLEM SYMPTOM CHECK THIS REMEDY No pow...

Page 13: ...e value 40 199 beat minute Device Classi cation Expected Lifetime Battery Powered Mode Internally Powered ME Equipment AC Adaptor Powered Mode Class II ME Equipment Temperature 5 C to 40 C Relative hu...

Page 14: ...s RF emissions CISPR 11 RF emissions CISPR 11 Group 1 Class B Class A Comply Compliance Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Emissions test Table 2 EMC...

Page 15: ...modi cations not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment NOTE This equipment has been tested and found to comply with the li...

Reviews: