The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments.

Warning: 

Don’t be near the active HF surgical equipment and the RF shielded room of an ME 

system for magnetic resonance imaging, where the intensity of EM disturbances is high.

Warning: 

Use of this equipment adjacent to or stacked with other equipment should be 

avoided because it could result in improper operation. If such use is necessary, this equipment 
and the other equipment should be observed to verify that they are operating normally.

Warning: 

Use of accessories, transducers and cables other than those specified or provided by 

the manufacturer of this equipment could result in increased electromagnetic emissions or 
decreased electromagnetic immunity of this equipment and result in improper operation.

Warning: 

Portable RF communications equipment (including peripherals such as antenna 

cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of 
the equipment CH-B2-11-AA including cables specified by the manufacturer. Otherwise, 
degradation of the performance of this equipment could result.

1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL 

PERFORMANCE with regard to electromagnetic disturbances for the excepted service life. 
2. Guidance and manufacturer’s declaration-electromagnetic emissions and Immunity.

Table 1 

Guidance and manufacturer’s declaration - electromagnetic emissions

RF emissions

CISPR 11 

RF emissions

CISPR 11 

Group 1 

Class B

Class A

Comply

Compliance

Harmonic emissions

IEC 61000-3-2 

Voltage fluctuations / flicker emissions

IEC 61000-3-3

Emissions test

Table 2 

EMC GUIDANCE

EMC GUIDANCE

Guidance and manufacturer’s declaration – electromagnetic Immunity 

NOTE  

U

is the a.c. mains voltage prior to application of the test level. 

±8 kV contact
±2 kV, ±4 kV, ±8 kV,  ±15 kV air 

±8 kV contact
±2 kV, ±4 kV, ±8 kV,  ±15 kV air 

Electrostatic
discharge (ESD) 
IEC 61000-4-2  

±2 kV for power supply lines

±1 kV signal input/output

100 kHz repetition frequency

±2 kV for power supply lines

NA 100 kHz repetition 

frequency

±0.5 kV, ±1 kV differential mode

±0.5 kV, ±1 kV, ±2 kV common mode

±0.5 kV, ±1 kV differential 

mode N/A

Electrical fast
transient/burst
IEC 61000-4-4   

Surge
IEC61000-4-5

Immunity Test

Compliance level

IEC 60601-1-2 
Test level

30 A/m
50 Hz / 60 Hz

30 A/m
50 Hz / 60 Hz

Power frequency 
magnetic field
IEC 61000-4-8

Voltage dips, short 
interruptions and 
voltage variations 
on power supply
 input lines 
IEC 61000-4-11

0% UT ; 0,5 cycle. At 0°, 45°, 90°, 

135°, 180°, 225°, 270° and 315°.

0% UT ; 1 cycle and 70% UT ;

25/30 cycles; Single phase: at 0°.

0% UT ; 250 / 300 cycle

0% UT  ; 0,5 cycle. At 0°, 45°, 

90°, 135°, 180°, 225°, 270° 

and 315°. 0% UT  ; 1 cycle and 

70% UT  ; 25/30 cycles; 

Single phase: at 0°. 0% UT ; 

250 / 300 cycle

Conduced RF
IEC61000-4-6

3 V

0,15 MHz – 80 MHz

6 V in ISM and amateur radio 

bands between 0,15 MHz and 

80 MHz 80% AM at 1 kHz

3 V

0,15 MHz – 80 MHz

6 V in ISM and amateur radio 

bands between 0,15 MHz and 

80 MHz

80% AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz

Radiated RF
IEC61000-4-3

26

25

Technical description:

Summary of Contents for Carbon Health CH-B1-11-AA

Page 1: ...1 0 Guangdong Transtek Medical Electronics Co Ltd Zone A No 105 Dongli Road Torch Development District 528437 Zhongshan Guangdong China User Manual Blood Pressure Monitor Bluetooth Connected Model CH...

Page 2: ...pressure and diastolic pressure What is the standard blood pressure classi cation Why does my blood pressure fluctuate throughout the day Why do i get a different blood pressure at home compared to t...

Page 3: ...e and also pulse rate This Pro Cellular Blood Pressure Monitor CH B2 11 AA is a digital monitor intended for use in measuring blood pressure and heartbeat rate with arm circumference ranging from 22 t...

Page 4: ...s component part lists descriptions and calibration procedures will be made available by the manufacturer or distributor This device is intended for indoor use This device is portable but it is not in...

Page 5: ...ation and air leakage testing at least at 50 mmHg and 200 mmHg Warning Do not use the device while under maintenance or being serviced Store your device cuff and adapter in a clean and dry place prote...

Page 6: ...ou nish the measurement 5V 1A Replace the batteries whenever the below happens The low battery symbol appears on the display When any button is pressed and nothing is displayed on the screen BATTERY C...

Page 7: ...oth and the Carbon Health App Make sure both are ON when pair up is proceeding 3 When the monitor is OFF press and hold the START STOP button unitl the symbol appears on the LCD and blink indicating p...

Page 8: ...o insert one nger between the cuff and your arm 2 3 cm 1 inch MEASUREMENT Interference may occur in the vicinity of equipment marked with the following symbol And CH B1 11 AA may interfer the vicinity...

Page 9: ...nd the oldest record will drops from the list after the measurement Caution You can either record your measurement results manually along with the time of measurement or transfer them electronically t...

Page 10: ...d avoid the sunshine Avoid intense shaking and collisions Use a damp cloth to remove dirt Avoid washing the cuff Avoid dusty and unstable temperature environments Within 20 minutes after taking a bath...

Page 11: ...e Please contact a physician if your measurement falls out of the range Please note that only a physician can tell whether your blood pressure value has reached a dangerous point The following chart i...

Page 12: ...for 4 5 minutes to calm down If the cuff is tied properly If the cuff is too tight or too loose If the cuff is tied on the upper arm If you feel anxious 22 21 PROBLEM SYMPTOM CHECK THIS REMEDY No pow...

Page 13: ...e value 40 199 beat minute Device Classi cation Expected Lifetime Battery Powered Mode Internally Powered ME Equipment AC Adaptor Powered Mode Class II ME Equipment Temperature 5 C to 40 C Relative hu...

Page 14: ...s RF emissions CISPR 11 RF emissions CISPR 11 Group 1 Class B Class A Comply Compliance Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Emissions test Table 2 EMC...

Page 15: ...modi cations not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment NOTE This equipment has been tested and found to comply with the li...

Reviews: