Frequency Measurement - Low Frequency Input
To set the low frequency measurement input, send the following command
MEASURE:INPUT:LOWFREQ<CR>
This will set the unit to Low Frequency measurement mode.
To receive the currently displayed measurement, read the response to the following command
READ?
The response will be returned in Hz with the same resolution as on the display
For example, if 1.0456 kHz is currently being measured with a gate time of 1s the response will
be
1045.60000
Frequency Measurement - High Frequency Input
To set the High Frequency measurement input, send the following command
MEASURE:INPUT:HIGHFREQ<CR>
This will set the unit to High Frequency measurement mode.
To receive the currently displayed measurement, read the response to the following command
READ?
The response will be returned in Hz with the same resolution as on the display
For example, if 1.0456 MHz is currently being measured with a gate time of 10s the response
will be
1045600000