Page 24
www.transmille.com
8600 Operation Manual - V1.0
Frequency Source - Low Frequency Sine
To set the sine wave low frequency output (Variable Frequency and level), send the following
command
SOURCE:LF:SINE<CR>
This command will enter the sine wave output mode.
To set the Impedance, send the following command
SOURCE:LF:SINE:IMPEDANCE {Data}<CR>
Where data is HIGH or LOW to set the correct impedance
To set the frequency, send the following command
SOURCE:LF:SINE:FREQ {Data}<CR>
Where data is the desired frequency in Hz
To set the Level, send the following command
SOURCE:LF:SINE:LEVEL {Data}<CR>
Where data is the desired level in volts
For Example, to set the output to High Input impedance, at 5 kHz with an output of 1V send
SOURCE:LF:SINE/SOURCE:LF:SINE:IMPEDANCE HIGH/
SOURCE:LF:SINE:FREQ 5000/SOURCE:LF:SINE:LEVEL 1<CR>